Published online Apr 18, 2020. doi: 10.5312/wjo.v11.i4.232
Peer-review started: October 29, 2019
First decision: December 12, 2019
Revised: December 19, 2019
Accepted: March 12, 2020
Article in press: March 12, 2020
Published online: April 18, 2020
Processing time: 167 Days and 6.5 Hours
Total hip replacement has become one of the most successful orthopaedic procedures. The length of the femoral stem constitutes one of the most important geometrical and mechanical features of the prosthesis. Several different implants are currently available but data are limited concerning the clinical results for some of these implants.
To report the short-term clinical and radiological results of a novel squared section, tapered design – with four conicity - short stem in total hip replacement.
This is a retrospective study of a prospectively collected data using of MINIMA® short stem in 61 consecutive patients with at least 1 year follow-up. The collected data included patients’ demographics, type of arthritis, bone morphology, perioperative data, clinical results using Harris Hip Score, EuroQol (EQ-5D), pain score and satisfaction rate, complications and radiological results.
Total 61 patients were included in our study with a mean age of 56 years of age (range 25-73 years). The majority of them (68.6%) were women, thirty seven patients (56.9%) were less than 60 years of age and almost half of patients (45.1%) suffered from secondary osteoarthritis (hip dysplasia, osteonecrosis, etc.). The mean time of follow-up examination was 33.4 mo (2.8 years) with a range of 12-57 months (1-4.8 years). In 35 patients (56.9%) the follow-up examination was more than 3 years. No major complications such as revision, periprosthetic fracture, dislocation or infection were presented. Re-admission 90 d postoperatively or laterwas deemed unnecessary for any reason regarding the operation. Respectively, the mean pain score, mean Harris hip score, and mean EQ-5D were improved from 6.3, 58.7 and 77.3 preoperatively to 0.1, 95.1, and 79.8 postoperatively. The Satisfaction rate at the final follow-up was 9.9 (SD 0.3, range 8.0-10.0). All stems were classified as stable bone ingrowth and no radiolucent lineswere revealed in any of the modified Gruens’ zone at the postoperative X-rays. Stem subsidence was within acceptable limits and the incidence of distal cortical hypertrophy was relatively low.
The clinical and radiological results concerning the MINIMA® short stem are excellent according to this first report of this specific design of the short femoral stems. Because of the small number of cases and short-term follow-up of this study, a longer follow up time and more patients’ enrollment is required.
Core tip: Total hip replacement has become one of the most successful orthopaedic procedures. The length of the femoral stem constitutes one of the most important geometrical and mechanical features of the prosthesis. Several different implants are currently available but data are limited concerning the clinical results for some of these implants. Our objective is to report the short-term results of a novel squared section, tapered design –with four conicity- short stem. According to our findings the clinical and radiological results were excellent in the short-term follow-up using the MINIMA® short stem in total hip replacement.