Basic Research
Copyright ©2005 Baishideng Publishing Group Inc.
World J Gastroenterol. Jan 28, 2005; 11(4): 521-524
Published online Jan 28, 2005. doi: 10.3748/wjg.v11.i4.521
Table 1 Range and precision of each sensor.
SensorRangePrecision
Pressure-60 to +200 mmHg1% FS
Temperature34 to 42 °C±0.2 °C
pH1 to 13±0.2
Table 2 Weight and dimension of each capsule.
ParameterPT capsulepH capsule
Diameter (mm)1010
Length (mm)21.124
Weight (gram)2.95.2
Table 3 Volunteers and CIB time.
Volunteer no.SexAge (yr)CIB time (h)
1M4225.78
2M5233.83
3M3428.22
4F5011.72
5M5519.07
6M5433.12
7M4918.77
Table 4 Pressure data.
PressureGauge data
(kPa)
(mmHg)
Up
Down
Up
Down
-8-60-59.6-60-60.4-60.7
-4-30-29.3-28.9-30.4-30.6
0001.2-0.4-0.4
5.33403840.43839.3
10.67807980.77979.8
16120121.2121.8120.8121.2
21.33160161.7162.5159.9160.9
26200200.8199.8199.8199.8
Table 5 Temperature data.
Temperature (°C)Gauge data
3434.02
3635.91
3838.03
4040.06
4242.11
Table 6 pH data.
pHGauge data
12345
44.44.34.34.34.4
6.866.76.76.86.76.7
9.189.29.39.39.29.3