Retrospective Study
Copyright ©The Author(s) 2023.
World J Clin Pediatr. Sep 9, 2023; 12(4): 230-236
Published online Sep 9, 2023. doi: 10.5409/wjcp.v12.i4.230
Table 1 Comparison of effective radiation dose on dual source computed tomography coronary angiography platforms in patients with Kawasaki disease
Ref.
Country
Number of cases
Platform
ECG triggering
Age
Radiation dose
Remarks
Duan et al[27], 2012China19DSCTProspectiveRange: 3 mo-5 yrED (mean ± SD): 0.36 ± 0.06 mSv
Kim and Goo[28], 2013Korea51DSCT: (1) 64-Slice: n = 49; and (2) 128-Slice: n = 2Retrospective in most of the patientsMean (range): 13.2 (1-24) yrED (mean ± SD): (1) 64-Slice: 2.6 ± 2.7 mSv; and (2) 128-Slice: 2.1 ± 0.6 mSvED was only 0.6 ± 0.5 mSV in 5 children < 2 yr of age, who underwent prospective ECG-triggering
Ghoshhajra et al[29], 2014United States52DSCT: (1) 64-Slice: n = 16; and (2) 128-Slice: n = 36ProspectiveRange: 0-18 yrMedian (IQR) ED: (1) 64-Slice: 2.9 (0.9-4.1) mSV; and (2) 128-Slice: 1.0 (0.6-2.0) mSVED significantly reduced compared to 16 and 64-Slice Scanner platforms
Kantarcı et al[30], 2019Turkey17128-Slice DSCTDetails NAMean (range): 3 yr (2 mo-11.3 yr)ED range: 1.2-4.3 mGy depending on the patient’s body weight
van Stijn et al[26], 2020Netherlands70DSCT: (1) 2 × 192-Slice: n = 56; and (2) Other CT scanners (64/128/320-Slice): n = 14ProspectiveMedian (range): 15.1 (0.5-59.5) yrMedian (range) ED: (1) 2 × 192-Slice: 1.5 (0.3-9.4) mSv; and (2) Other CT scanner: 3.8 (1.7-20.0) mSvThis is the only study on 3rd generation DSCT platform
Borhanuddin et al[31], 2022Malaysia5264-slice DSCTRetrospectiveMedian (range): 5 (1-18) yrMedian (range) ED: 0.81 (0.4-5.8)
Present study, ChandigarhIndia85128-Slice DSCTProspectiveMedian (range): 5 yr (2 mo-11 yr)Median (IQR) ED: 0.83 (0.68-1.01)Largest study on 128-Slice platform which provides data that CTCA can be performed in sub-millisievert doses
Table 2 Conversion factors for chest in different age groups at 80 kV according to recent International Commission on Radiological Protection recommendations (20)
S.No.
Age group
Tube voltage (kV)
Conversion factor
1< 1 yr800.0823
2> 1 - < 5 yr800.0525
3> 5 - < 10 yr800.0344
4≥ 10 yr800.0248
Table 3 Effective computed tomography radiation exposure according to age in adaptive prospective electrocardiography-triggered sequence computed tomography coronary angiography on 128-dual source computed tomography platform
S.No.
Age group
Mean DLP (mGy.cm)
Mean effective radiation dose (mSv)
10-1 yr (n = 10)9.19 ± 2.210.63 ± 0.16
21-5 yr (n = 29)18.82 ± 8.480.85 ± 0.41
35-10 yr (n = 38)31.79 ± 14.141.04 ± 0.37
4> 10 yr (n = 8)55.67 ± 19.391.38 ± 0.48