World J Anesthesiol. Mar 27, 2014; 3(1): 119-123
Published online Mar 27, 2014. doi: 10.5313/wja.v3.i1.119
Published online Mar 27, 2014. doi: 10.5313/wja.v3.i1.119
Hour | #1 | #2 | #3 | #4 | #5 | |
On-Q | 1 | 19 | 19.4 | 18.6 | 17.9 | 18.4 |
2 | 18 | 18.4 | 18.4 | 16.8 | 18 | |
3 | 17.6 | 18.4 | 17.9 | 16.1 | 17.5 | |
4 | 17.8 | 18 | 16.7 | 16.4 | 17.5 | |
5 | 17.6 | 18.2 | 15.8 | 15.4 | 17.5 | |
6 | 17.5 | 18.2 | 14.9 | 16.4 | 16.9 | |
7 | 18.6 | 18.2 | 14.9 | 16.2 | 16.9 | |
8 | 19.3 | 19.8 | 14.7 | 16.1 | 16.8 | |
9 | 18.8 | 18.4 | 14.9 | 14.9 | 16.6 | |
10 | 18.5 | 19 | 15.1 | 14.8 | 16.6 | |
Baxter | 1 | 10.8 | 10.4 | 10.8 | 10.4 | 9.8 |
2 | 10.1 | 10 | 10.6 | 10.3 | 10 | |
3 | 10 | 9.8 | 10.7 | 9.9 | 9.8 | |
4 | 9.8 | 9.9 | 10.5 | 10 | 9.7 | |
5 | 9.7 | 9.5 | 10.1 | 10 | 10 | |
6 | 9.8 | 9.8 | 10.5 | 10 | 9.8 | |
7 | 9.6 | 9.6 | 10.4 | 9.8 | 9.9 | |
8 | 10.5 | 9.8 | 10.4 | 10 | 9.8 | |
9 | 10 | 10.1 | 10.3 | 10 | 9.6 | |
10 | 10 | 10.1 | 10.7 | 10.1 | 9.6 | |
Ambu | 1 | 9.5 | 9.2 | 10 | 8.5 | 10.2 |
2 | 9.8 | 9.2 | 9.8 | 8.6 | 9.9 | |
3 | 9.8 | 9.2 | 9.5 | 8.6 | 9.8 | |
4 | 9.6 | 9.2 | 9.8 | 8.8 | 10.2 | |
5 | 9.7 | 9.2 | 9.8 | 8.6 | 9.9 | |
6 | 9.6 | 9.2 | 9.8 | 8.9 | 9.9 | |
7 | 9.7 | 9.2 | 9.7 | 8.8 | 10.4 | |
8 | 9.7 | 9.2 | 9.9 | 8.8 | 10.2 | |
9 | 9.4 | 9.1 | 9.4 | 8.5 | 9.9 | |
10 | 9.6 | 9.2 | 10 | 8.7 | 9.8 |
Pump | n | Mean | StdDev | Std Err | P value1 |
On-Q | 10 | 17.3 | 0.62 | 0.19 | < 0.0001 |
Baxter | 10 | 10.1 | 0.17 | 0.05 | < 0.0001 |
AMBU | 10 | 9.5 | 0.09 | 0.03 | < 0.0001 |
- Citation: LeRiger M, Bhalla T, Martin D, Bettesworth J, Tobias JD. Comparison of flow rate accuracy and consistency between the on-Q, baxter, and ambu pain infusion devices. World J Anesthesiol 2014; 3(1): 119-123