Table 1 Spinal deformity angles
Mean | SD | P value (t - test) | |
Pre-distraction PA cobb | 53.8 | 19.9 | |
Post-distraction PA cobb | 50.2 | 20 | 0.003 |
Pre-distraction Lat cobb | 46.2 | 13.7 | |
Post-distraction Lat cobb | 41.8 | 13.5 | 0.006 |
Rods distraction (mm) | 19.2 | 4.4 |
Table 2 Force/distraction values
Distraction (mm) | Min. load (N) | Mean load (N) | Max. load (N) |
1 (min) | -32 (min) | 133 (min) | 311 (min) |
2 | 30 | 203 | 394 |
3 | 141 | 247 | 447 |
4 | 167 | 282 | 465 |
5 | 172 | 308 | 447 |
6 | 201 | 354 | 552 (max) |
7 | 246 | 364 | 428 |
8 | 222 | 354 | 440 |
9 | 303 | 402 | 461 |
10 | 342 | 430 | 497 |
11 | 370 | 404 | 445 |
12 (max) | 472 (max) | 485 (max) | 498 |
- Citation: Teli M, Grava G, Solomon V, Andreoletti G, Grismondi E, Meswania J. Measurement of forces generated during distraction of growing-rods in early onset scoliosis. World J Orthop 2012; 3(2): 15-19