Copyright
©2012 Baishideng Publishing Group Co.
World J Radiol. Jan 28, 2012; 4(1): 21-28
Published online Jan 28, 2012. doi: 10.4329/wjr.v4.i1.21
Published online Jan 28, 2012. doi: 10.4329/wjr.v4.i1.21
Baseline (200 mAs) | Unpro (150 mAs) | Pro (150 mAs) | Unpro (100 mAs) | Pro (100 mAs) | Unpro (50 mAs) | Pro (50 mAs) | |
No. of lesions | 31 | 31 | 31 | 31 | 31 | 31 | 31 |
Lesion size (< 1 cm) | 28 | 28 | 28 | 28 | 28 | 28 | 28 |
Lesion conspicuity | 1 | 1 | 1 | 1 | 1 | 1 | 1 |
Image noise | 2 | 3 | 2 | 4 | 2 | 5 | 4 |
Visibility of small structures | 3 | 3 | 3 | 4 | 3 | 4 | 4 |
Image contrast | 3 | 3 | 3 | 3 | 3 | 3 | 3 |
Diagnostic confidence | 1 | 1 | 1 | 2 | 1 | 4 | 3 |
Baseline (200 mAs) | Unpro (150 mAs) | Pro (150 mAs) | Unpro (100 mAs) | Pro (100 mAs) | Unpro (50 mAs) | Pro (50 mAs) | |
Liver | |||||||
Objective noise | 15.5 ± 3.9 | 18.2 ± 4.9 | 14.4 ± 4.0 | 21.9 ± 5.6 | 15.7 ± 4.0 | 30.4 ± 9.2 | 18.8 ± 5.0 |
HU values | 70.9 ± 15.0 | 70.5 ± 16.9 | 70.6 ± 16.0 | 71.2 ± 16.0 | 71.0 ± 16.0 | 70.1 ± 16.0 | 70.1 ± 16.0 |
Abdominal Aorta | |||||||
Objective noise | 17.4 ± 4.2 | 20.5 ± 5.4 | 16.0 ± 3.8 | 24.8 ± 6.2 | 18.5 ± 4.2 | 34.3 ± 9.3 | 22.0 ± 6.0 |
HU values | 80.5 ± 29.0 | 81.6 ± 37.0 | 81.5 ± 31.0 | 82.0 ± 33.7 | 82.1 ± 33.0 | 80.3 ± 31.0 | 80.0 ± 31.0 |
Unprocessed images | Processed images | |||
Acceptable noise | Unacceptable noise | Acceptable noise | Unacceptable noise | |
150 mAs | 14/19 | 5/19 | 19/19 | 0/19 |
Weight (kg) | 74.5 ± 14.8 | 95.2 ± 11.1 | 79.9 ± 16.5 | - |
Transverse diameter (cm) | 34.2 ± 4.3 | 42.9 ± 4.5 | 36.5 ± 5.8 | - |
100 mAs | 4/19 | 15/19 | 17/19 | 2/19 |
Weight (kg) | 77.1 ± 4.0 | 80.7 ± 18.6 | 76.9 ± 14.6 | 106.0 ± 4.3 |
Transverse diameter (cm) | 32.5 ± 3.5 | 37.6 ± 5.9 | 35.7 ± 5.6 | 43.5 ± 0.7 |
50 mAs | 0/19 | 19/19 | 5/19 | 14/19 |
Weight (kg) | - | 79.9 ± 16.5 | 66.2 ± 13.0 | 91.4 ± 15.1 |
Transverse diameter (cm) | - | 36.5 ± 5.8 | 33.0 ± 3.4 | 40.6 ± 6.1 |
- Citation: Singh S, Kalra MK, Sung MK, Back A, Blake MA. Radiation dose reduction with application of non-linear adaptive filters for abdominal CT. World J Radiol 2012; 4(1): 21-28
- URL: https://www.wjgnet.com/1949-8470/full/v4/i1/21.htm
- DOI: https://dx.doi.org/10.4329/wjr.v4.i1.21