Table 1 Single straight pipe phantom: intra-operator variability of semi-quantitative perfusion parameters following five 0.02/0.05/0.08 mL bolus injections of SonoVue®
PI | TPI | Slope of the WI | FWHM | AUC | AUWI | AUWO | |
0.02 mL/CV (%) | 6.03 | 9.78 | 9.57 | 6.43 | 9.40 | 9.97 | 11.11 |
0.05 mL/CV (%) | 9.27 | 10.14 | 8.28 | 13.24 | 5.23 | 12.76 | 6.84 |
0.08 mL/CV (%) | 20.87 | 10.10 | 23.52 | 17.43 | 5.87 | 23.72 | 2.94 |
Table 2 Three intertwined pipe phantom: intra-operator variability of semi-quantitative perfusion parameters following five 0.03/0.06 mL bolus injections of SonoVue®
PI | TPI | Slope of the WI | FWHM | AUC | AUWI | AUWO | |
0.03 mL/CV (%) | 3.16 | 7.69 | 9.65 | 5.56 | 2.33 | 8.55 | 5.32 |
0.06 mL/CV (%) | 7.85 | 8.60 | 11.62 | 7.53 | 6.73 | 9.52 | 6.12 |
Table 3 Dialyzer: intra-operator variability of semi-quantitative perfusion parameters following five 0.10 mL bolus injections of SonoVue®
PI | TPI | Slope of the WI | FWHM | AUC | AUWI | AUWO | |
0.10 mL/CV (%) | 9.66 | 8.11 | 10.27 | 19.11 | 9.31 | 13.89 | 13.36 |
Table 4 In vivo studies: intra-operator variability of semi-quantitative perfusion parameters following three 0.02/0.05/0.1 mL bolus injections of SonoVue®
Mouse | PI | TPI | Slope of the WI | FWHM | AUC | AUWI | AUWO | ||
0.02 mL/ 0.05 mL/ 0.1 mL/ CV (%) | Chemical anaesthesia | 1 | 18.99 | 14.04 | 23.85 | 28.87 | 15.79 | 18.43 | 15.46 |
2 | 27.60 | 8.33 | 25.75 | 18.90 | 9.06 | 28.39 | 8.12 | ||
3 | 11.39 | 4.41 | 1.48 | 18.44 | 10.76 | 14.73 | 13.58 | ||
4 | 27.88 | 5.68 | 28.82 | 29.97 | 11.21 | 26.22 | 12.42 | ||
Gaseous anaesthesia | 5 | 1.90 | 17.63 | 10.09 | 24.96 | 12.32 | 12.43 | 12.79 |
Table 5 Acoustic properties of different media
Density (kg/m3) | Velocity (m/s) | Attenuation coefficient (dB/cm at 1 MHz) | Impedance (MRayl) | |
Blood | 1057 | 1575 | 0.18 | 1.61 |
Water | 1000 | 1480 | 0.00221 | 1.51 (50°C) |