Gauthier M, Leguerney I, Thalmensi J, Chebil M, Parisot S, Peronneau P, Roche A, Lassau N. Estimation of intra-operator variability in perfusion parameter measurements using DCE-US. World J Radiol 2011; 3(3): 70-81 [PMID: 21512654 DOI: 10.4329/wjr.v3.i3.70]
Corresponding Author of This Article
Marianne Gauthier, MSc, IR4M, UMR 8081, CNRS, Paris-Sud 11 Univ, Gustave Roussy Institute, Villejuif 94805, France. marianne.gauthier@igr.fr
Article-Type of This Article
Original Article
Open-Access Policy of This Article
This article is an open-access article which was selected by an in-house editor and fully peer-reviewed by external reviewers. It is distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/
World J Radiol. Mar 28, 2011; 3(3): 70-81 Published online Mar 28, 2011. doi: 10.4329/wjr.v3.i3.70
Table 1 Single straight pipe phantom: intra-operator variability of semi-quantitative perfusion parameters following five 0.02/0.05/0.08 mL bolus injections of SonoVue®
PI
TPI
Slope of the WI
FWHM
AUC
AUWI
AUWO
0.02 mL/CV (%)
6.03
9.78
9.57
6.43
9.40
9.97
11.11
0.05 mL/CV (%)
9.27
10.14
8.28
13.24
5.23
12.76
6.84
0.08 mL/CV (%)
20.87
10.10
23.52
17.43
5.87
23.72
2.94
Table 2 Three intertwined pipe phantom: intra-operator variability of semi-quantitative perfusion parameters following five 0.03/0.06 mL bolus injections of SonoVue®
PI
TPI
Slope of the WI
FWHM
AUC
AUWI
AUWO
0.03 mL/CV (%)
3.16
7.69
9.65
5.56
2.33
8.55
5.32
0.06 mL/CV (%)
7.85
8.60
11.62
7.53
6.73
9.52
6.12
Table 3 Dialyzer: intra-operator variability of semi-quantitative perfusion parameters following five 0.10 mL bolus injections of SonoVue®
PI
TPI
Slope of the WI
FWHM
AUC
AUWI
AUWO
0.10 mL/CV (%)
9.66
8.11
10.27
19.11
9.31
13.89
13.36
Table 4 In vivo studies: intra-operator variability of semi-quantitative perfusion parameters following three 0.02/0.05/0.1 mL bolus injections of SonoVue®
Citation: Gauthier M, Leguerney I, Thalmensi J, Chebil M, Parisot S, Peronneau P, Roche A, Lassau N. Estimation of intra-operator variability in perfusion parameter measurements using DCE-US. World J Radiol 2011; 3(3): 70-81