Table 1 Sonazoid-enhanced US at a low mechanical index (MI)
AuthorsInjection volume of SonazoidMI in late phaseScan starting time in late phase
Moriyasu et al[33]0.015 mL/kg0.3-0.510 min afterinjection
Korenaga et al[42]0.01 mL/kg0.2-0.330 min after injection
Hatanaka et al[39]0.01 mL/kg0.2At least 10 min after injection
Maruyama et al[44]0.0075 mL/kg0.24-0.35-10 min after injection
Numata et al[37]0.2 mL/body0.2At least 5 min after injection
Table 2 Setting conditions of Sonazoid-enhanced US at a low and a high MI
ParametersLow MIHigh MI
Transducer3.5 or 4.0-MHz convex transducer
Imaging modeCoded phase inversion (A), Contrast pulse sequence (B), Pulse inversion harmonic (C), Amplitude modulation (D)Coded harmonic angio (A)
MI0.12-0.3 (average 0.2)0.7-1.0 (average 0.8)
Frame rate7-13 frames2 -13 frames
Focus positionBottom of the liver (A, D), Bottom of the lesion (B, C)Bottom of the lesion (A)
Table 3 Comparison between Sonazoid-enhanced US images at low and high MI for evaluation of liver lesion vascularity
ParametersLow MIHigh MI
Vascular phase images
Tumor vessel imageGood<Very good
Tumor enhancement imageGood<Very good
Tumor enhancement image of the lesion located in deep portion or high echoic lesionsPoor<<Good
Real time image (frame rate)Good (high frame rate)>Good to poor (high to low frame rate)
Late phase images
Scan starting timeAt least 10 min<At least 5 min
Perfusion defect image of the lesionGood<Very good
Perfusion defect images of the lesion located in deep portion or high echoic lesionPoor<<Good
Real-time image (frame rate)Very good (high frame rate)>>Poor (low frame rate)
Examination timeMore than 10 min<More than 5 min
Injection volume of SonazoidHalf or the volume recommended<Quarter or third of the volume recommended