Kawai N, Minamiguchi H, Sato M, Nakai M, Sanda H, Tanaka T, Ikoma A, Nakata K, Shirai S, Sonomura T. Evaluation of vascular puncture needles with specific modifications for enhanced ultrasound visibility: In vitro study. World J Radiol 2012; 4(6): 273-277 [PMID: 22778880 DOI: 10.4329/wjr.v4.i6.273]
Corresponding Author of This Article
Morio Sato, MD, Professor, Department of Radiology, Wakayama Medical University, 811-1 Kimiidera, Wakayama-shi, Wakayama 641-8510, Japan. morisato@wakayama-med.ac.jp
Article-Type of This Article
Brief Article
Open-Access Policy of This Article
This article is an open-access article which was selected by an in-house editor and fully peer-reviewed by external reviewers. It is distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/
Nobuyuki Kawai, Hiroki Minamiguchi, Morio Sato, Motoki Nakai, Hiroki Sanda, Takami Tanaka, Akira Ikoma, Kouhei Nakata, Shintaro Shirai, Tetsuo Sonomura, Department of Radiology, Wakayama Medical University, 811-1 Kimiidera, Wakayama-shi, Wakayama 641-8510, Japan
ORCID number: $[AuthorORCIDs]
Author contributions: Kawai N and Minamiguchi H contributed equally to this work; Kawai N, Nakai M and Sanda H designed the research; Ikoma A, Tanaka T and Nakata K performed the research; Nakata K and Ikoma A contributed new reagents/analytic tools; Shirai S and Sonomura T analyzed the data; Sato M and Kawai N wrote the paper.
Correspondence to: Morio Sato, MD, Professor, Department of Radiology, Wakayama Medical University, 811-1 Kimiidera, Wakayama-shi, Wakayama 641-8510, Japan. morisato@wakayama-med.ac.jp
Telephone: +81-73-4410604 Fax: +81-73-4443110
Received: February 29, 2012 Revised: June 14, 2012 Accepted: June 21, 2012 Published online: June 28, 2012
Abstract
AIM: To determine which modification to a vascular puncture needle results in increased visualization during ultrasound (US)-guided vascular puncture.
METHODS: We evaluated US images of a phantom made of degassed gelatin and each of the following four modified versions of a commercially available vascular puncture needle (18 G): re-cut needle, dimple needle, rough-surface needle (rough over the sections of needle located 3-6 mm from the tip), and a needle with four side holes (side holes covered by the sheath). An unmodified commercially available puncture needle was used as a control. Five interventional radiologists evaluated image quality according to the following classification grade: I, invisible; II, poor; III, moderate; IV, good; V, excellent.
RESULTS: The highest score for needle visualization was obtained for the needle with four side holes. The re-cut needle scored the same as the control. Multiple comparisons were conducted using overall evaluation scores among the commercially available needle, dimple needle, rough-surface needle (3-6 mm), and the needle with four side holes. A significantly higher score was obtained for the needle with four side holes (P < 0.05/6).
CONCLUSION: The needle with four side holes was prominently visualized and gained a significantly higher score (compared with the other needles) in a phantom evaluation.
Citation: Kawai N, Minamiguchi H, Sato M, Nakai M, Sanda H, Tanaka T, Ikoma A, Nakata K, Shirai S, Sonomura T. Evaluation of vascular puncture needles with specific modifications for enhanced ultrasound visibility: In vitro study. World J Radiol 2012; 4(6): 273-277
A rise in the number of cancer patients worldwide has resulted in an increased frequency of central venous (CV) port placement for chemotherapy or high caloric infusion[1,2]. Interventional radiologists are commonly requested to perform venous puncture of the jugular, subclavian, and femoral veins[3-12]. Furthermore, venous puncture of the popliteal vein is necessary for treatment of deep venous thrombosis[13]. Ultrasound (US) guidance enables the puncture to be performed precisely and quickly, with a low incidence of complications[1-8,10,14,15]. However, complications such as arterial puncture, pneumothorax, and hemothorax still occur following US-guided puncture, probably because of unclear needle visualization[1-3,8,11,16]. Previous studies have described devices for gauging the puncture angle and needle depth, and the use of four-dimensional US for improving needle visualization[17,18]. Biopsy needles and nerve-block needles with a specific coating or type of cut at the needle tip are commercially available for improving US needle visualization[19-26]. Nevertheless, no vascular needle has been devised specifically for this purpose. We consider that a specific modification to the vascular needle itself would improve needle visualization and thus increase puncture accuracy.
The purpose of this study is to create four types of needle modification for improving US visualization, and to identify the needle modification that is most suitable for increasing needle visualization during US-guided puncture, in testing using an in vitro phantom.
MATERIALS AND METHODS
We used a commercially available vascular needle (Medikit, Tokyo, Japan) composed of an inner needle (20 gauge stainless steel, 0.9 mm in outer diameter and 0.6 mm in hollow diameter) and an outer sheath (18 gauge, 1.3 mm, Teflon). We created the following modified versions of the commercially available puncture needle: a re-cut needle, a dimple needle, a file-like rough-surface needle, and a needle with four side holes (Figure 1). The commercially available needle (unmodified) was used as a control (Figure 2). The re-cut needle was made by re-cutting the dorsal aspect of the needle tip (Figure 3). The shaft of the rough-surface needle was filed to a rough surface at a distance of 3-6 mm from the tip (Figure 4). The dimple needle had numerous small dimples (0.08 mm in diameter) added to the shaft at a distance of 5-8 mm from the tip (Figure 5). To produce the needle with side holes, we drilled four side holes of 0.4 mm diameter into the shaft at 4 mm from the tip (Figure 6). The modified regions of the dimple, rough surface, and side hole needles were covered by the sheath.
Figure 1 Schematic drawings of a commercially available puncture needle and four newly devised vascular puncture needles (18 gauge).
Top to bottom: re-cut needle, dimple needle, rough-surface needle (rough over the section of needle located 3-6 mm from the tip), and needle with four side holes.
Figure 3 Vascular needle with re-cutting of the dorsal aspect of the tip.
A: Needle tip; B: Corresponding ultrasound image. The arrow indicates the needle tip.
Figure 4 Vascular needle with a rough-filed surface at 3-6 mm from the tip of the needle.
A: Needle tip, shaft; B: Corresponding ultrasound image, with the rough surface indicated by the arrow.
Figure 5 Vascular needle with dimples placed at 5-8 mm from the tip (dimple needle).
A: Needle tip, shaft; B: Corresponding ultrasound image, with the dimples indicated by the arrowhead and the needle tip indicated by the arrow.
Figure 6 Vascular needle with four side holes, each 0.
4 mm in diameter, drilled at 4 mm from the tip of the needle. A: Needle tip, shaft. The side holes are covered by the outer sheath. B: Corresponding ultrasound image. The arrow indicates the needle tip with side holes.
We performed an in vitro experimental study using a Xario US machine (Toshiba Medical, Tokyo, Japan) with an 11 MHz linear-array probe and a phantom made of degassed gelatin. The probe-to-surface angle was approximately 30°. Under US guidance, the phantom was punctured by each of the five needles, to a depth of approximately 1.5 cm from the surface of the gelatin. We obtained five images of needle puncture for each needle. The total of 25 images (5 images × 5 needles) was piled and arranged at random. Five interventional radiologists with more than 2 years experience in CV port placement checked the image quality of the needle and evaluated its practical utility for vascular puncture according to a classification grade. The grades were as follows: I, invisible; II, poor; III, moderate; IV, good; V, excellent.
Statistical analysis
We performed multiple comparisons among the commercially available needle, dimple needle, rough-surface needle, and the needle with four side holes (the re-cut needle was excluded because similar scores were obtained between this needle and the commercially available needle). The Mann-Whitney test was used to compare categorical variables of grade scores given by the radiologists among the commercially available needle, dimple needle, rough-surface needle, and the needle with four side holes. Bonferroni’s method was used to adjust for multiple (6 times) comparisons post-test. For the Mann-Whitney test and Bonferroni’s correction, P < 0.05 and P < 0.05/6 were considered to indicate significant difference, respectively.
RESULTS
The scores for needle visualization are listed in Table 1. The highest scores for needle visualization were obtained for the needle with four side holes, followed by the dimple needle, the rough-surface needle, and the re-cut needle. The needle with the re-cut modification yielded the same score as the control. Multiple comparisons were conducted using the control needle and the three needles with the modifications whose scores were significant higher than the control, and the significantly highest score was obtained for the needle with four side holes (P < 0.05/6).
Table 1 Ultrasound visibility of variously modified vascular needles.
The ease of performing needle puncture of the subclavian vein depends on the width of the vein and the thickness of subcutaneous fat[4,6,11]. A subclavian vein that is thin due to dehydration is sometimes difficult to puncture, even if US is used[4]. Thick subcutaneous fat tissue pushes the subcutaneous vein deep, causing a long needle puncture route and generating increased echo from the thick fat, causing poor visualization of the puncture needle[4,11]. In deep vein thrombus, poor visualization of the puncture needle during needle puncture of the popliteal vein is caused by the elevated echogenicity of the thrombus in the vessels and thick limb edema, with subcutaneous fat pushing the popliteal vein deep[13]. Therefore, modifications to improve visualization of the vascular puncture needle would improve safety during these procedures[17,20-26].
Of the four modified needles assessed in the present study, the re-cut modification showed no significant increase in needle visibility compared with the control. Of the three modifications that increased visibility, the needle with four side holes was visualized most prominently on US, followed by the dimpled needle and the rough-surface needle. The difference in visualization appears to depend on the size of the modification: the minute grooves of the rough surface had the least effect on echogenicity, while the large-sized modification of the 0.35-mm side holes had the greatest effect on echogenicity, followed by the 0.08-mm dimple modification. We consider that the larger the size of the modification, the greater the echo reflection generated, and that the adherence of air to the four side holes enhanced echogenicity.
A weakness of the present study is that it is limited to an in vitro study, rather than the clinical situation. A number of clinical cases should be warranted for the possible use of the needle with four side holes. The degree of visibility of the puncture needle in human fat tissue could be greater or less than that in the phantom. However, no fundamental data exist regarding the visibility of puncture needles. The present data will provide a basis for considering modifications to puncture needles. It would be possible to combine the features of side holes and rough surface in a puncture needle, and this appears to be a valid subject for future development.
In conclusion, we created and tested four modified needles for vessel puncture with the aim of enhancing their visibility under US. Among the devices, needle tip visualization was most prominent in the needle with side holes, and its visualization score was significantly higher in evaluation using a phantom.
COMMENTS
Background
Ultrasound (US) guidance enables vascular puncture to be performed precisely and quickly, with a low incidence of complications. However, complications such as arterial puncture, pneumothorax, and hemothorax still occur following US-guided puncture, probably because of unclear needle visualization.
Research frontiers
No vascular needle has been devised specifically for clear needle visualization. The authors consider that a specific modification to the vascular needle itself would improve needle visualization and thus increase puncture accuracy.
Innovations and breakthroughs
The authors created four types of needle modification for improving US visualization, and to identify the needle modification that is most suitable for increasing needle visualization during US-guided puncture. Four modified versions of a commercially available vascular puncture needle (18 G) were evaluated: re-cut needle, dimple needle, rough-surface needle (rough over the sections of needle located 3-6 mm from the tip), and a needle with four side holes (side holes covered by the sheath). The highest score for needle visualization was obtained for the needle with four side holes. The needle with four side holes was prominently visualized and gained a significantly higher score (compared with the other needles) in a phantom evaluation.
Applications
The needle with four side holes could be applied in cases with a thin subclavian vein due to hydration, thick fat causing a long needle puncture route and poor visualization of the puncture needle, and would improve safety during these procedures.
Peer review
The paper under consideration is very new and should be accepted for publication in the wider interest as a preliminary study.
Footnotes
Peer reviewer: Satish Kumar Bhargava, Professor, Department of Radiology, University College of Medical Sciences, 201, Fancy Apartment, 19, Vasundhara Enclave, Delhi-96, Delhi 110096, India
Gebauer B, El-Sheik M, Vogt M, Wagner HJ. Combined ultrasound and fluoroscopy guided port catheter implantation--high success and low complication rate.Eur J Radiol. 2009;69:517-522.
[PubMed] [DOI][Cited in This Article: ]
Orihashi K, Imai K, Sato K, Hamamoto M, Okada K, Sueda T. Extrathoracic subclavian venipuncture under ultrasound guidance.Circ J. 2005;69:1111-1115.
[PubMed] [DOI][Cited in This Article: ]
Verghese ST, McGill WA, Patel RI, Sell JE, Midgley FM, Ruttimann UE. Ultrasound-guided internal jugular venous cannulation in infants: a prospective comparison with the traditional palpation method.Anesthesiology. 1999;91:71-77.
[PubMed] [DOI][Cited in This Article: ]
Koroglu M, Demir M, Koroglu BK, Sezer MT, Akhan O, Yildiz H, Yavuz L, Baykal B, Oyar O. Percutaneous placement of central venous catheters: comparing the anatomical landmark method with the radiologically guided technique for central venous catheterization through the internal jugular vein in emergent hemodialysis patients.Acta Radiol. 2006;47:43-47.
[PubMed] [DOI][Cited in This Article: ]
Galinski M, Catineau J, Tazarourte K, Dardel N, Bertrand P, Adnet F, Lapostolle F. Comparison of historical anatomic landmarks vs. ultrasound guidance for the selection of a needle insertion site for jugular central venous access.Resuscitation. 2012;83:e113-e114.
[PubMed] [DOI][Cited in This Article: ][Cited by in Crossref: 1][Cited by in F6Publishing: 4][Article Influence: 0.3][Reference Citation Analysis (0)]
Chumnanvej S, Wetchagama N, Kounsongtham V. Accuracy of needle-tip localization by ultrasound guidance lumbar selective nerve root block: a prospective clinical study of 78 lumbar nerve roots block.J Med Assoc Thai. 2011;94:1451-1456.
[PubMed] [DOI][Cited in This Article: ]
Castillo D, McEwen DS, Young L, Kirkpatrick J. Micropuncture needles combined with ultrasound guidance for unusual central venous cannulation: desperate times call for desperate measures--a new trick for old anesthesiologists.Anesth Analg. 2012;114:634-637.
[PubMed] [DOI][Cited in This Article: ][Cited by in Crossref: 4][Cited by in F6Publishing: 6][Article Influence: 0.5][Reference Citation Analysis (0)]
Deam RK, Kluger R, Barrington MJ, McCutcheon CA. Investigation of a new echogenic needle for use with ultrasound peripheral nerve blocks.Anaesth Intensive Care. 2007;35:582-586.
[PubMed] [DOI][Cited in This Article: ]
Schafhalter-Zoppoth I, McCulloch CE, Gray AT. Ultrasound visibility of needles used for regional nerve block: an in vitro study.Reg Anesth Pain Med. 2004;29:480-488.
[PubMed] [DOI][Cited in This Article: ]
Tavakoli M, Kellar EJ, Nassiri D, Joseph AE. A novel polymeric coating for enhanced ultrasound visibility of medical devices.Med Device Technol. 2006;17:8-10, 12.
[PubMed] [DOI][Cited in This Article: ]