Detection | Threshold (bpm) | Beats to detect | Therapies |
VF | 250 | 18 out of 24 | 30 J × 6 |
FVT | 200 | 18 out of 24 | 1 × burst, 30 J × 5 |
VT | 150 | 16 | 2 × burst, 1 × ramp, 30 J × 3 |
Detection | Threshold (bpm) | Beats to detect | Therapies |
VF | 250 | 18 out of 24 | 30 J/35 J × 6 |
FVT | 182 | 18 out of 24 | 1 × Burst, 30 J/35 J × 6 |
VT | 167 | 32 | Off |
- Citation: De Maria E, Giacopelli D, Borghi A, Modonesi L, Cappelli S. Antitachycardia pacing programming in implantable cardioverter defibrillator: A systematic review. World J Cardiol 2017; 9(5): 429-436