World J Cardiol. Oct 26, 2013; 5(10): 382-386
Published online Oct 26, 2013. doi: 10.4330/wjc.v5.i10.382
Published online Oct 26, 2013. doi: 10.4330/wjc.v5.i10.382
FBP | IR | P value | |
Age (yr) | 55.6 ± 9.1 | 56.0 ± 10.1 | 0.67 |
Male | 85 (85) | 80 (80) | 0.35 |
Body mass index (kg/m2) | 27.2 ± 2.7 | 26.3 ± 3.4 | 0.03 |
Body mass index ≥ 30 | 15 (15) | 13 (13) | 0.68 |
Heart rate (bpm) | 58.3 ± 7.0 | 58.2 ± 6.4 | 0.88 |
Acquisition technique | |||
Prospective (axial) | 84 (84) | 82 (82) | 0.71 |
Retrospective (helical) | 16 (16) | 18 (18) | |
Tube voltage (kV) | 119.0 ± 4.4 | 109.0 ± 10.4 | < 0.0001 |
Percent 80-100 kV | 5 (5) | 54 (54) | < 0.0001 |
mAs in prospective | 203.1 ± 15.4 | 195.7 ± 26.8 | < 0.0001 |
mAs in helical | 943.2 ± 119.5 | 870.1 ± 122.8 | < 0.0001 |
Radiation dose (mSv) | |||
Total | 4.6 ± 3.0 | 2.8 ± 1.4 | < 0.0001 |
Prospective (axial) | 3.4 ± 2.4 | 2.4 ± 0.7 | < 0.0001 |
Retrospective (helical) | 10.3 ± 3.9 | 5.2 ± 1.6 | < 0.0001 |
Image quality | |||
Attenuation level (HU) | 461.7 ± 51.9 | 462.2 ± 51.2 | 0.54 |
Image noise (HU) | 37.8 ± 1.4 | 38.2 ± 2.4 | 0.20 |
Signal to noise ratio | 12.2 ± 1.4 | 12.1 ± 1.4 | 0.28 |
Coronary assessment (%) | 91.7 ± 4.0 | 92.5 ± 2.8 | 0.12 |
- Citation: Carrascosa P, Rodriguez-Granillo GA, Capuñay C, Deviggiano A. Low-dose CT coronary angiography using iterative reconstruction with a 256-slice CT scanner. World J Cardiol 2013; 5(10): 382-386
- URL: https://www.wjgnet.com/1949-8462/full/v5/i10/382.htm
- DOI: https://dx.doi.org/10.4330/wjc.v5.i10.382