Minireviews
Copyright ©The Author(s) 2022.
World J Cardiol. Aug 26, 2022; 14(8): 446-453
Published online Aug 26, 2022. doi: 10.4330/wjc.v14.i8.446
Table 1 Internal and external device interaction with permanent pacemakers / defibrillators

Erquo et al[8], 2018
Pfeffer et al[9], 2016
Obwegeser et al[10], 2001
Shenoy et al[15], 2017
Singh et al[14], 2018
Schimpf et al[12], 2003
Internal/ External DeviceSt. Jude; CRT-D Boston Scientific; S-ICDMedtronic dual chamber ICDDurata dual chamber ICDBoston Scientific; S-ICD; Medtronic dual chamber ICD
EMI SourceLVAD (pump exchange)LVAD placementThalamic DBSTENS unit (unreported location)TENS unit on neck, back, axillaT11 SCS
Adverse InteractionVentricular oversensing leading to inhibition of pacingOversensing leading to 31 ICD shocksNoneEMI leading to ICD shockEMI leading to ICD shockNone
Therapeutic interventionTurning off low-frequency attenuation filterTransvenous ICD placementN/AAvoiding TENS unitsAvoiding TENS unitsN/A
Table 2 Device interaction on pacemaker and defibrillators
Type of device
Pacemaker
Defibrillators
LVADs++
SCS--
DBS--
TENS++
Table 3 Suggested cardiac implantable electronic devices peri-procedural management
Procedure
CIED (PPM)
CIED (ICD)
EndoscopyAsynchronousR or Magnet
APCAsynchronousR or Magnet
ENBbb
ElectrocauteryAsynchronousR or Magnet