Erquo et al[8], 2018 | Pfeffer et al[9], 2016 | Obwegeser et al[10], 2001 | Shenoy et al[15], 2017 | Singh et al[14], 2018 | Schimpf et al[12], 2003 | |
Internal/ External Device | St. Jude; CRT-D | Boston Scientific; S-ICD | Medtronic dual chamber ICD | Durata dual chamber ICD | Boston Scientific; S-ICD; | Medtronic dual chamber ICD |
EMI Source | LVAD (pump exchange) | LVAD placement | Thalamic DBS | TENS unit (unreported location) | TENS unit on neck, back, axilla | T11 SCS |
Adverse Interaction | Ventricular oversensing leading to inhibition of pacing | Oversensing leading to 31 ICD shocks | None | EMI leading to ICD shock | EMI leading to ICD shock | None |
Therapeutic intervention | Turning off low-frequency attenuation filter | Transvenous ICD placement | N/A | Avoiding TENS units | Avoiding TENS units | N/A |
Type of device | Pacemaker | Defibrillators |
LVADs | + | + |
SCS | - | - |
DBS | - | - |
TENS | + | + |
Procedure | CIED (PPM) | CIED (ICD) |
Endoscopy | Asynchronous | R or Magnet |
APC | Asynchronous | R or Magnet |
ENB | b | b |
Electrocautery | Asynchronous | R or Magnet |
- Citation: Barmore W, Patel H, Voong C, Tarallo C, Calkins Jr JB. Effects of medically generated electromagnetic interference from medical devices on cardiac implantable electronic devices: A review. World J Cardiol 2022; 14(8): 446-453