Review Open Access
Copyright ©2011 Baishideng Publishing Group Co., Limited. All rights reserved.
World J Cardiol. Mar 26, 2011; 3(3): 84-92
Published online Mar 26, 2011. doi: 10.4330/wjc.v3.i3.84
Advantages and disadvantages of biodegradable platforms in drug eluting stents
Agustina Rodriguez-Granillo, Center for Advanced Biotechnology and Medicine, Rutgers University, New Jersey, NJ 08854, United States
Bibiana Rubilar, Staff Cardiac Unit, Otamendi Hospital, Buenos Aires C1024AAA, Argentina
Gaston Rodriguez-Granillo, Intravascular Imaging, Cardiac Unit, Otamendi Hospital, Buenos Aires C1024AAA, Argentina
Alfredo E Rodriguez, Cardiac Unit, Otamendi Hospital, Cardiovascular Research Center (CECI), Buenos Aires C1024AAA, Argentina
Correspondence to: Alfredo E Rodriguez, MD, PhD, FACC, FSCAI, Head, Director, Cardiac Unit, Otamendi Hospital, Cardiovascular Research Center (CECI), Callao 1441 4B Capital Federal, Buenos Aires C1024AAA, Argentina. rodrigueza@sanatorio-otamendi.com.ar
Telephone: +54-11-49648721 Fax: +54-11-49629012
Received: January 8, 2011
Revised: February 14, 2011
Accepted: February 21, 2011
Published online: March 26, 2011

Abstract

Coronary angioplasty with drug-eluting stent (DES) implantation is currently the most common stent procedure worldwide. Since the introduction of DES, coronary restenosis as well as the incidence of target vessel and target lesion revascularization have been significantly reduced. However, the incidence of very late stent thrombosis beyond the first year after stent deployment has more commonly been linked to DES than to bare-metal stent (BMS) implantation. Several factors have been associated with very late stent thrombosis after DES implantation, such as delayed healing, inflammation, stent mal-apposition and endothelial dysfunction. Some of these adverse events were associated with the presence of durable polymers, which were essential to allow the elution of the immunosuppressive drug in the first DES designs. The introduction of erodable polymers in DES technology has provided the potential to complete the degradation of the polymer simultaneously or immediately after the release of the immunosuppressive drug, after which a BMS remains in place. Several DES designs with biodegradable (BIO) polymers have been introduced in preclinical and clinical studies, including randomized trials. In this review, we analyze the clinical results from 6 observational and randomized studies with BIO polymers and discuss advantages and disadvantages of this new technology.

Key Words: Polymers; Drug eluting stents; Biodegradable polymers; Stents; Thrombosis; Restenosis



INTRODUCTION

A significant reduction in coronary restenosis rates has been observed with the introduction of drug-eluting stent (DES) technology during percutaneous coronary interventions (PCI)[1-6]. During these years, we have also learned that some adverse effects, although rarely present, are more frequently associated with DES implantation[7-14]; some of them can be linked to durable polymers, which were continually present in the first DES designs[1-6]. Delayed healing, endothelial dysfunction, chronic arterial wall inflammation and late-acquired stent mal-apposition are more frequently linked with DES implantation[9-15]. All of these can increase the incidence of very late stent thrombosis that, although an uncommon event, was more frequently reported after DES implantation[16-21]. Delayed loss of anti-restenotic efficacy was also reported with the first DES designs[22,23]. Chronic arterial wall inflammation and endothelial dysfunction may be associated with the increased rate of target vessel revascularization (TVR) at a late stage, which has been found particularly in patients with complex lesions including those with diabetes[24,25]. All of the above underscore the importance of this topic, especially after reports of increased rate of endothelial dysfunction after DES implantation as compared with bare-metal stent (BMS) implantation.

FIRST DES DESIGNS

Three main components were necessary to achieve a stable release of the drug in the first DES generation: the stent platform to scaffold the vessel, the polymer to deliver the immunosuppressive agent and the drug to inhibit neointimal growth.

Initially, sirolimus-eluting stents (SES; Cypher™, Cordis Co., Warren, NJ, USA) and paclitaxel-eluting stents (PES; Taxus™, Boston Scientific Co., Natick, MA, USA) were designed using permanent polymers such as poly(ethylene-co-vinyl acetate), poly(n-butyl methacrylate) and poly(styrene-b-isobutylene-b-styrene), which allowed controlled elution of the immunosuppressive agent. The SES design consists of a stainless steel platform coated with a permanent polymer containing sirolimus 140 μg/cm2, 80% of which is released within 30 d. The PES design is also composed of a stainless steel platform with a permanent polymer coating combined with 1 μg/mm2 paclitaxel; 10% of the paclitaxel is released within 2 wk after stent deployment, although 90% of it remains in the polymer forever[24,26].

The presence of permanent polymers in the vessel arterial wall adds an additional factor that influences local responses and may alter processes involved in neointimal formation. Each polymer provokes a distinctive inflammatory response in animals, such as giant cell infiltration around the stent struts, and a progressive granulomatous and eosinophilic reaction[23,27], which increase beyond the first year. These data support the perception that durable polymers in DES technology may provoke chronic inflammation and decreased efficacy.

BIODEGRADABLE POLYMERS IN DES TECHNOLOGY

The use of biodegradable (BIO) polymers, as opposed to durable polymers, in coronary stent technology has the advantages of a complete elution of drugs and a reduced inflammatory response, with the potential for decreasing the risk of late complications such as stent strut uncovering, mal-apposition, endothelial dysfunction and thrombosis[5,7,9,10,28,29]. BIO polymers allow the complete release and elution of the immunosuppressive agent after degradation of the polymer[26,29]. Therefore, long-term antiplatelet therapy would not be required after the polymer completely disappeared. The most common BIO polymers are composed of polylactic acid (PLA), polyglycolide and poly(D,L-lactic-co-glycolic acid) (PLGA), which are completely metabolized into the body (breaking down into monomers, water and carbon dioxide) after fulfilling their purpose.

Several new stents with fully BIO polymers have been introduced using a variety of anti-proliferative agents such as sirolimus, tacrolimus, biolimus and paclitaxel. The safety and efficacy of these devices have been assessed clinically in first in man (FIM) and observational studies[30-32]. For example, in the Paclitaxel In Stent Controlled Elution Study, the pharmacokinetics of the DES and not dose of the immunosuppressive agent appears to be associated with neointimal suppression and clinical outcome[30]. However, an excess of late loss with a high rate of clinical angiographic restenosis and also a lack of reduction in stent thrombosis were reported in many of these FIM studies. A high inflammatory reaction due to major particle debris as a result of coating degradation, which was not simultaneous with drug release, was a major limitation for many of the first DES designs with BIO polymers. Therefore, to the best of our knowledge, only 6 DES with erodable polymers have randomized clinical data with enough patients to justify their introduction in clinical practice.

In the following paragraphs we will review and summarize the main findings from published randomized data of the Limus Eluted from a Durable vs Erodable Stent Coating (LEADERS)[28], NOBORI[33,34], Individualized Drug Eluting Stent System to Abrogate Restenosis (ISAR)-TEST-3[35] and TEST-4[36,37] with sirolimus (rapamycin), PAINT (percutaneous intervention with BIO-polymer based paclitaxel-eluting or sirolimus-eluting vs bare stents for de novo coronary lesions)[38] and EUCATAX[39] trials. Study and stent design of each trial is described in Table 1.

Table 1 Comparison between published trials of biodegradable eluting stents.
NamePolymerStent designDrugDrug per stent length (μg/mm2)Polymer degradationDrug release
ISAR-TEST-3[35]PLA316L stainless steel microporus stentSirolimus4.86-9 wk28 d (95%)
ISAR TEST-4[36]PLA316L stainless-steel microporus stentSirolimus4.86-9 wk28 d (95%)
NOBORI 1[34]PLAStainless-steel S-stentBiolimus15.69-12 moTwo phases: immediately after stent implantation; sustained drug release over 9-12 mo
NOBORI CORE[33]PLAStainless-steel S-stentBiolimus15.69-12 moTwo phases: immediately after stent implantation; sustained drug release over 9-12 mo
LEADERS[28]PLAFlexible stainless- steel stentBiolimus15.66-9 mo6-9 mo
PAINT[38]PLA+316L stainless metallic platformPaclitaxel and Sirolimus6.4 (PES)7 mo9-11 d (50%)
PLGA6.6 (SES)38 d (90%)
48 d (100%)
EUCATAX[39]PLGAStainless steel open cell with glycocalix layerPaclitaxel11 to 436-8 wk6-8 wk
LEADERS TRIAL

The LEADERS trial[28] is the largest randomized trial with BIO polymer-coated stents. The study compared a PLA polymer loaded with Biolimus (Biolimus-eluting stent; BioMatrix Flex, Biosensors Inc, Newport Beach, CA, USA) vs a Cypher platform (SES). The BIO polymer was applied to the stent’s abluminal surface only. After an initial burst of 40% of drug elution, complete drug release and polymer degradation was achieved over a period of 6 to 9 mo.

The LEADERS trial enrolled 1707 randomized patients, 807 included in the BIO polymer (BioMatrix Flex) and 850 in the durable polymer (Cypher) DES arms. The study included a large proportion of patients with acute coronary syndromes (55%) including ST elevation myocardial infarction (STEMI), multiple vessel disease (24%), previous PCI (36%) and vessel size < 2.75 mm (68%). At 9 mo of follow-up, all clinical endpoints met the criteria of non-inferiority for the BioMatrix-Flex compared to Cypher, including the amount of late loss in the follow-up angiogram.

The incidence of all definitions of stent thrombosis was also similar between both DES designs (3.6% in BioMatrix-Flex and 3.3% in Cypher), although percent of uncovered (3.6% vs 39% in BioMatrix-Flex and Cypher, respectively, P = 0.005) or mal-apposed (0.3% vs 6.7% in BioMatrix-Flex and Cypher, respectively, P = 0.04) stent struts were significant higher in the Cypher stent arm. At 2 years of follow-up, the LEADERS trial also showed a superior outcome with BioMatrix in patients with STEMI as testified by the rates of major adverse cardiovascular events (MACCE; 8.1% for BioMatrix-Flex vs 19.3% for Cypher, P < 0.01); the incidence of stent thrombosis in this cohort of patients was significantly lower with BioMatrix-Flex compared with Cypher (2.6% vs 8.4%, respectively, P < 0.05).

In the short-term follow-up, there was a high incidence of non-STEMI in patients allocated to the BioMatrix-Flex polymer (5.4%), and a high incidence of stent thrombosis in patients with STEMI allocated to the SES arm.

NOBORI TRIAL

The Nobori stent (Terumo Co., Tokyo, Japan) uses a similar drug-polymer combination (Biolimus/PLA) as the one in BioMatrix in the LEADERS trial. In this trial[33,34], 243 patients were randomized in a 2:1 ratio between Biolimus with BIO polymer stents (Nobori) and paclitaxel with durable polymer stents (Taxus).

At 9 mo of follow-up, the use of a DES with BIO polymer compared with the Taxus Liberte DES significantly reduced the amount of late loss and angiographic restenosis. Furthermore, although not powered to detect clinical differences, the incidence of target lesion revascularization (TLR) and TVR were also significantly better with the Nobori stent design. Remarkably, stent thrombosis was not seen in the erodable polymer arm compared with 4.4% in the Taxus Liberte arm. There was a small sample size, a short-term outcome and a high rate of stent thrombosis in the Taxus Liberte arm.

ISAR-TEST-3 AND -TEST-4 TRIALS
ISAR-TEST-3 trial

The ISAR-TEST-3 trial[35] enrolled 605 patients randomized to a BIO polymer stent loaded with sirolimus, a sirolimus polymer-free stent and a sirolimus with permanent polymer stent (Cypher; Cordis, Florida, USA). The BIO polymer was completely absorbed within 6 to 9 wk after stent deployment, whereas 100% of sirolimus was released within the first 30 d. The main finding of this study was that the BIO polymer stent was not inferior to the Cypher stent in safety and efficacy, whereas the polymer-free stent was inferior in terms of efficacy to the other 2 DES designs.

This study had a small sample size, a short-term outcome and, in this DES stent design, the BIO polymer remains in place after the drug is completely eluted, therefore inflammatory reactions by the polymer itself cannot be excluded.

ISAR-TEST-4 trial

The ISAR-TEST-4 trial[36] enrolled 2603 patients randomized to a BIO polymer DES (1299) or a permanent polymer DES (1304). In the latest group, 652 patients were treated with a Cypher stent and 652 with Xience V (Abbott Vascular, Abbott Park, IL, USA). At 1 year of follow-up, there were no differences in angiographic and clinical endpoints among patients treated with a BIO or permanent polymer, and the stent clearly met the non-inferiority test in both cases.

Two years follow-up of this trial was recently presented[37], and a sustained equivalence in the incidence of safety/efficacy end points between BIO and permanent polymer DES designs was seen. The incidence of stent thrombosis was similar in both study arms.

This study had a short-term outcome and the same concerns regarding the BIO polymer DES design described above for ISAR-TEST-3.

PAINT TRIAL

The PAINT trial[38] compared 2 DES with the same BIO polymer but with a different immunosuppressive drug (paclitaxel or sirolimus) vs a BMS design; 274 patients were randomly assigned to paclitaxel with BIO polymer (n = 111), sirolimus with BIO polymer (n = 106) and BMS (n = 57). All stents had the same laser cut stainless steel platform. Both paclitaxel and sirolimus were released in approximately 48 d, whereas complete polymer degradation occurred only after 7 mo.

Both DES designs had less late loss and TVR compared with the BMS, whereas SES had a lower late loss but similar 1-year clinical outcome compared with PES. All-cause death, MI and stent thrombosis were similar in the 3 groups.

The study had a small sample size, short period of follow-up, greater amount of in stent late loss with both BIO polymers in comparison with historical studies with the same drugs but a permanent polymer. Although the study did not show any significant differences in stent thrombosis rate among the different groups, this event occurred in 1.9% of each DES design compared with zero in the BMS arm. Finally, taking into consideration that release of the drug and degradation of the polymer was not simultaneous, similar concerns with this stent design can be applied to those described in the ISAR-TEST studies.

EUCATAX TRIAL

The rationale and purpose of the EUCATAX trial[39] was to compare the efficacy and safety of a new PES dual coated with a BIO polymer and glycocalyx against an equivalent BMS (Eucatech AG, Reinhelfeden; Germany). A FIM study was previously conducted[40].

The PES is a stainless steel open cell (strut thickness 85 μm) modular design with 3 connecting fins per modulo. The double coating includes a BIO polymer as the platform for paclitaxel elution and a glycocalyx to increase hemocompatibility. The glycocalyx layer is a symmetric coating that uses camouflage nanotechnology. The BIO polymer is PLGA, which forms an asymmetric coating with a thickness of 2.5 μm on the luminal side and 5 μm on the abluminal side. Paclitaxel is loaded into the polymer, at a concentration of 11 to 43 μg depending on the stent length. The camouflage nanocoating[41] is coated with hemo-heparin, which is a polymer-analogous modified heparin that lacks an active anticoagulation effect due to removal of the sulfate groups. On top of this hemoheparin coat, the bio-absorbable polyester polymer PLGA serves as the carrier of the paclitaxel. In this stent design, degradation of the polymer occurred simultaneously with the elution of the drug at 6 to 8 wk after deployment. Therefore, according to the manufacturers, neither the drug nor the polymer remained in place (Figure 1).

Figure 1
Figure 1 EUCATAX design and characteristics. PLGA: Poly(D,L-lactic-co-glycolic acid).

The study included 422 patients (9.1% of those screened) and randomized 211 patients to the PES arm and 211 to the BMS arm (Figure 2). The population included diabetes in 23.5%, a reference vessel diameter size < 2.75 mm in 60%, multi-vessel disease in 60%, and acute coronary syndrome in 60%.

Figure 2
Figure 2 EUCATAX randomized trial design. Modified form Rodriguez et al[39]. Pts: Patients; PCI: Percutaneous coronary interventions; PES: Paclitaxel eluting stent; BMS: Bare-metal stent.

Cumulative clinical events at 18.3 ± 7.3 mo are shown in Figure 3. Cumulative cardiac events such as death, cardiac death, MI and stroke were similar, although PES showed a lower incidence of TLR and TVR driving a significant reduction in target vessel failure (TVF) and MACCE compared with the BMS design, and both were the major end points of the study. The incidence of any stent thrombosis was 1.4% in the PES group and 1.9% in the BMS group. Interestingly, beyond 1 year, no patient in the PES arm suffered stent thrombosis.

Figure 3
Figure 3 Results at 18 mo. EUCATAX trial. AMI: Acute myocardial infarction; MACCE: Major adverse cardiovascular events; TVF: Target vessel failure; TLR: Target lesion revascularisation; TVR: Target vessel revascularisation; PES: Paclitaxel-eluting stents; BMS: Bare-metal stent.

Baseline and follow-up angiographic findings are shown in Table 2. Follow-up angiography was performed in all 150 patients scheduled. In segment late luminal loss was 0.50 mm in the PES group and 0.94 mm in the BMS group (P = 0.001). The binary restenosis rate was 13.2% (13/98 lesions) in the PES arm and 34% (30/88 lesions) in the BMS arm (P < 0.001). Intravascular ultrasound showed no difference in late incomplete stent mal-apposition between groups (Table 2), although the incidence of late incomplete mal-apposition in the proximal segment of the stent was significantly in favor of the PES group (P = 0.015).

Table 2 Quantitative coronary analysis for both groups in the EUCATAX trial n (%).
PESBMSP value
Baseline QCA analysisn = 169n = 153
Reference diameter (mm)2.75 ± 0.52.85 ± 0.50.086
Minimal luminal diameter (mm)0.86 ± 0.40.85 ± 0.50.780
Lesion length (mm)16.2 ± 6.115.6 ± 6.30.410
Stent diameter (mm)21.7 ± 5.620.0 ± 4.80.160
Stent size (mm)2.96 ± 0.42.93 ± 0.50.780
Immediately Post PCI QCA analysis
Reference diameter (mm)2.91 ± 0.442.96 ± 0.430.340
Minimal luminal diameter (mm)2.68 ± 0.422.72 ± 0.430.400
Follow up QCA analysisn = 98n = 88
Reference diameter (mm)2.75 ± 0.482.75 ± 0.360.990
Minimal luminal diameter (mm)2.16 ± 0.511.81 ± 0.750.007
Stenosis diameter (%)27.4 ± 29.839.6 ± 23.90.005
Acute gain1.82 ± 0.471.87 ± 0.620.450
Net gain1.30 ± 0.490.93 ± 0.630.002
Late loss (in-stent)0.52 ± 0.590.94 ± 0.700.002
Late loss (in-segment)0.50 ± 0.560.91 ± 0.0690.001
Angiographic restenosis13 (13.2)31 (35.2)0.001
Follow up intravascular ultrasound analysisn = 45n = 37
Stent length (mm)21.7 ± 5.620.0 ± 4.80.160
Stent size (mm)2.96 ± 0.42.93 ± 0.50.780
Incomplete stent apposition5 (11.1)9 (24.3)0.150
Proximal segment1 (2.2)8 (21.6)0.015
Body segment2 (4.4)1 (2.7)1.000
Distal segment2 (4.4)0 (0.0)0.500

The study had a small sample size, a short-term outcome and a higher amount of late loss.

SUMMARY AND PERSPECTIVE

Since the introduction of the first DES designs, we have a strong clinical evidence for their significant benefits in terms of reduction of angiographic and clinical restenosis, which has been the Achilles heel of PCI during the past 30 years. However, although uncommon, we have also identified the potential deleterious effects of late and very late stent thrombosis associated with the implantation of these devices. Therefore, we clearly understand the complex process of designing the ideal DES, in which a combination of safety and efficacy should be the main goal.

Currently, we clearly recognize the advantages and disadvantages of the first DES designs in comparison with BMS, either in short- or long-term outcomes. However, little is known about the new DES generation in comparison with either BMS or the first DES designs. In the current review we report the short-term outcomes of new DES designs with BIO polymers either with respect to the first DES or to BMS designs. The results from these trials are presented in Table 3. Theoretically, BIO polymers have the advantage of complete degradation of the polymer together with the immunosuppressive agent that was loaded on it, after which only a BMS remains in place. Thus, all side effects related to durable polymers would be avoided or minimized with this type of coating. Requirements for dual antiplatelet therapy over a long period, mandatory with SES and PES with durable polymers, would now be necessary only within the period before polymer degradation. Long-term antiplatelet therapy is one of the major limitations for Cypher and Taxus implantation, especially in older patients or in patients with concomitant non-cardiac vascular or non-vascular illness requiring surgery. Consequently, there is plenty scope to improve the safety profile of the first DES generation; however, are these new stents with BIO polymers the answers to our concerns

Table 3 Comparison between published trials of biodegradable eluting stents.
NameStent designCardiac deathCardiac death or MIMITVRTLR
LEADERS[28]Biomatrix2.16.75.87.86.5
Cypher2.76.64.69.97.4
Nobori0.0-4.77.10.0
NOBORI[34]Taxus0.0-8.614.32.9
ISAR-TEST-3[35]Biodegradable polymer stent2.02.51.5-5.9
Permanent polymer sirolimus2.03.52.0-7.9
Polymer free sirolimus2.04.02.5-12.9
ISAR-TEST-4[36]Biodegradable polymer2.86.34.313.78.8
Control13.26.24.113.99.4
EUCATAX[39]PES1.94.72.88.26.1
BMS1.94.32.415.012.6

If we look at the results from the randomized LEADERS[28] and NOBORI[34] trials, with stent designs that share an identical polymer and drug, we do not see any advantage in terms of efficacy and safety in comparison with the old SES design. Even though the LEADERS study met the criteria for non-inferiority for the BioMatrix-Flex stent at 12 mo of follow-up, if we exclude the subgroup of patients with STEMI, we do not see any advantages in terms of safety in relation to the SES with a durable polymer (Cypher). Furthermore, the rate of non-STEMI reported in this trial with a BIO polymer stent (BioMatrix-Flex) design appears to be higher than we would expect. In addition, analysis from the subgroup of patients with STEMI had the bias of a high number of stent thromboses in the Cypher arm (over 8%), which was never reported in any randomized study in patients with STEMI with this DES design in the first year of follow-up[42].

The ISAR-TEST-4 trial[36] also reported a randomized head-to-head comparison between a BIO polymer SES vs 2 different durable coating DES designs, Cypher and Xience V. The ISAR BIO DES design had sirolimus completely released within the first 29 d, although the polymer disappeared between 6 and 9 wk after stent deployment. Therefore, the polymer in the stent remained in place around 1 mo after the release of the drug. For this reason, we cannot discard some adverse effects related to the polymer, free of drug, during that time. One- and two-year outcomes of this positive non-inferiority trial did not demonstrate any safety or efficacy advantage compared with the durable polymer arms, and the incidence of stent thrombosis was similar in all groups. Also, a late luminal catch up loss phenomenon between 8 mo and 2 years was reported in the ISAR-TEST-4 trial with the BIO polymer stent, a finding which was also commonly reported after implantation of SES (Cypher) with a durable polymer[37]. Taking into account that the drug and the polymer did not simultaneously disappear, an inflammatory response to coating breakdown cannot be discarded with this stent design.

In the PAINT trial[38], in both BIO stent designs the polymer remained in place for several months after the drug was completely eluted; therefore, an inflammatory response in response to the polymer itself and during the degradation process should be strongly considered.

In the EUCATAX trial[39], the BIO DES design allowed the polymer and the immunosuppressive drug to disappear simultaneously in the first 6 to 8 wk after deployment of the stent; beyond that time, a BMS with camouflage nanocoating remained in place. The camouflage nanocoating design has been linked with the safety outcome in promoting stent re-endothelialization and seems very useful in patients with a high-risk thrombotic profile such as STEMI or who underwent non-cardiac surgery soon after stent deployment. Interestingly, in a previous study with this kind of coating, intravascular ultrasound in patients with STEMI did not detect late acquired stent mal-apposition during follow-up angiography[43].

If we compare the EUCATAX with the LEADERS trial, which share similar clinical and angiographic inclusion criteria, excluding a significant lower late loss in favor of the LEADERS stent designs, the dual DES coating of EUCATAX showed similar rates of cardiac events including TLR, TVR and MI, although in the EUCATAX trial a trend of lower rates of MI and stent thrombosis were seen (Table 4). However, the down side of the last study was the large amount of late loss determined in the late angiography study with the EUCATAX stent design, which was higher than we expected (Table 2). Thus, we cannot exclude an inflammation process as a result of a crack in the PLGA coating, suggesting a possible breakdown of the polymer during the degradation process. We have to recognize that the degradation process of the BIO polymers is not always uniform; in poorly vascularized areas this process is likely to be slow, whereas in inflammatory areas it may be accelerated; consequently, if the drug elutes faster than the polymer, the advantage of the BIO polymer disappears.

Table 4 Comparison of baseline characteristics and follow up angiographic and clinical results from the randomized LEADERS[28] and EUCATAX[39] trials.
Patients characteristicsBioMatrixCypherEucaTaxP values
No. of patients857850211
Age (yr)65 ± 1165 ± 1163.8 ± 10.20.32
Male gender75.075.083.40.62
Hypertension74.073.064.00.46
Diabetes mellitus26.023.023.20.49
Hypercholesterolemia65.068.056.90.36
Smoking24.025.021.30.68
Previous MI32.033.020.40.02
Previous PCI36.037.035.50.93
Multi vessel disease37.032.055.0< 0.001
Clinical presentation
Acute coronary syndrome55.056.059.70.80
Lesions per patient
> 1 lesion29.022.026.10.09
Small vessels168.069.060.30.45
Procedural characteristics
Stents per lesion1.3 ± 0.71.3 ± 0.71.36 ± 0.50.38
Stent length per lesion (mm)24.7 ± 15.524.6 ± 14.821.7 ± 5.60.006
Angiographic follow-up
In-stent late loss0.08 ± 0.40.15 ± 0.40.52 ± 0.6< 0.001
Stent thrombosis2
Overall stent thrombosis3.63.31.40.28
Definite ST
0-30 d1.61.60.50.43
> 30 d-12 mo0.40.50.90.52
0 d-12 mo2.02.01.40.82
Efficacy endpoints at 12 mo
Any TLR6.57.46.10.63
Any TVR7.89.98.20.23
Safety endpoints at 12 mo
All causes of death3.23.32.40.80
Cardiac death2.12.71.90.66
Myocardial infarction5.84.62.80.11
Cardiac death or MI6.76.64.70.46
CONCLUSION

Introduction of completely BIO instead of durable polymers has the potential to avoid or minimize some of the side effects related to the first DES designs. One year follow-up results from these randomized trials have demonstrated similar safety/efficacy profiles with this new DES technology using BIO polymers when compared with durable polymer designs (LEADERS and ISAR trials). However, these similarities do not mean any superiority in terms of reduction of stent thrombosis, the Damocles sword of the first DES technology. Equivalency in efficacy requires longer follow-up assessment.

Dual coating technology using an antithrombotic layer behind the PLGA coating is promising in terms of safety, although its value in terms of efficacy is questionable and needs further assessment. Consequently, the complex process of designing a DES with BIO polymers remains a challenge.

Footnotes

Peer reviewers: Hiroki Teragawa, MD, PhD, Department of Cadiovascular Medicine, Hiroshima General Hospital of West Japan Railway Company, 3-1-36 Futabanosato, Higashi-ku, Hiroshima 732-0052, Japan; Tommaso Gori, MD, PhD, II Medizinische Klinik, Universitätsmedizin der Johannes Gutenberg Universitats Mainz, 55131 Mainz, Germany; Masamichi Takano, MD, PhD, Cardiovascular Center, Chiba-Hokusoh Hospital, Nippon Medical School, 1715 Kamakari, Imba, Chiba, 270-1694, Japan

S- Editor Cheng JX L- Editor Cant MR E- Editor Zheng XM

References
1.  Sousa JE, Costa MA, Abizaid AC, Rensing BJ, Abizaid AS, Tanajura LF, Kozuma K, Van Langenhove G, Sousa AG, Falotico R. Sustained suppression of neointimal proliferation by sirolimus-eluting stents: one-year angiographic and intravascular ultrasound follow-up. Circulation. 2001;104:2007-2011.  [PubMed]  [DOI]  [Cited in This Article: ]
2.  Morice MC, Serruys PW, Sousa JE, Fajadet J, Ban Hayashi E, Perin M, Colombo A, Schuler G, Barragan P, Guagliumi G. A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization. N Engl J Med. 2002;346:1773-1780.  [PubMed]  [DOI]  [Cited in This Article: ]
3.  Moses JW, Leon MB, Popma JJ, Fitzgerald PJ, Holmes DR, O'Shaughnessy C, Caputo RP, Kereiakes DJ, Williams DO, Teirstein PS. Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery. N Engl J Med. 2003;349:1315-1323.  [PubMed]  [DOI]  [Cited in This Article: ]
4.  Grube E, Silber S, Hauptmann KE, Mueller R, Buellesfeld L, Gerckens U, Russell ME. TAXUS I: six- and twelve-month results from a randomized, double-blind trial on a slow-release paclitaxel-eluting stent for de novo coronary lesions. Circulation. 2003;107:38-42.  [PubMed]  [DOI]  [Cited in This Article: ]
5.  Colombo A, Drzewiecki J, Banning A, Grube E, Hauptmann K, Silber S, Dudek D, Fort S, Schiele F, Zmudka K. Randomized study to assess the effectiveness of slow- and moderate-release polymer-based paclitaxel-eluting stents for coronary artery lesions. Circulation. 2003;108:788-794.  [PubMed]  [DOI]  [Cited in This Article: ]
6.  Stone GW, Ellis SG, Cox DA, Hermiller J, O'Shaughnessy C, Mann JT, Turco M, Caputo R, Bergin P, Greenberg J. A polymer-based, paclitaxel-eluting stent in patients with coronary artery disease. N Engl J Med. 2004;350:221-231.  [PubMed]  [DOI]  [Cited in This Article: ]
7.  Virmani R, Guagliumi G, Farb A, Musumeci G, Grieco N, Motta T, Mihalcsik L, Tespili M, Valsecchi O, Kolodgie FD. Localized hypersensitivity and late coronary thrombosis secondary to a sirolimus-eluting stent: should we be cautious. Circulation. 2004;109:701-705.  [PubMed]  [DOI]  [Cited in This Article: ]
8.  Togni M, Windecker S, Cocchia R, Wenaweser P, Cook S, Billinger M, Meier B, Hess OM. Sirolimus-eluting stents associated with paradoxic coronary vasoconstriction. J Am Coll Cardiol. 2005;46:231-236.  [PubMed]  [DOI]  [Cited in This Article: ]
9.  Hong MK, Mintz GS, Lee CW, Park DW, Park KM, Lee BK, Kim YH, Song JM, Han KH, Kang DH. Late stent malapposition after drug-eluting stent implantation: an intravascular ultrasound analysis with long-term follow-up. Circulation. 2006;113:414-419.  [PubMed]  [DOI]  [Cited in This Article: ]
10.  Kotani J, Awata M, Nanto S, Uematsu M, Oshima F, Minamiguchi H, Mintz GS, Nagata S. Incomplete neointimal coverage of sirolimus-eluting stents: angioscopic findings. J Am Coll Cardiol. 2006;47:2108-2111.  [PubMed]  [DOI]  [Cited in This Article: ]
11.  Joner M, Finn AV, Farb A, Mont EK, Kolodgie FD, Ladich E, Kutys R, Skorija K, Gold HK, Virmani R. Pathology of drug-eluting stents in humans: delayed healing and late thrombotic risk. J Am Coll Cardiol. 2006;48:193-202.  [PubMed]  [DOI]  [Cited in This Article: ]
12.  Meier P, Zbinden R, Togni M, Wenaweser P, Windecker S, Meier B, Seiler C. Coronary collateral function long after drug-eluting stent implantation. J Am Coll Cardiol. 2007;49:15-20.  [PubMed]  [DOI]  [Cited in This Article: ]
13.  Finn AV, Joner M, Nakazawa G, Kolodgie F, Newell J, John MC, Gold HK, Virmani R. Pathological correlates of late drug-eluting stent thrombosis: strut coverage as a marker of endothelialization. Circulation. 2007;115:2435-2441.  [PubMed]  [DOI]  [Cited in This Article: ]
14.  Nakazawa G, Finn AV, Joner M, Ladich E, Kutys R, Mont EK, Gold HK, Burke AP, Kolodgie FD, Virmani R. Delayed arterial healing and increased late stent thrombosis at culprit sites after drug-eluting stent placement for acute myocardial infarction patients: an autopsy study. Circulation. 2008;118:1138-1145.  [PubMed]  [DOI]  [Cited in This Article: ]
15.  Cook S, Wenaweser P, Togni M, Billinger M, Morger C, Seiler C, Vogel R, Hess O, Meier B, Windecker S. Incomplete stent apposition and very late stent thrombosis after drug-eluting stent implantation. Circulation. 2007;115:2426-2434.  [PubMed]  [DOI]  [Cited in This Article: ]
16.  McFadden EP, Stabile E, Regar E, Cheneau E, Ong AT, Kinnaird T, Suddath WO, Weissman NJ, Torguson R, Kent KM. Late thrombosis in drug-eluting coronary stents after discontinuation of antiplatelet therapy. Lancet. 2004;364:1519-1521.  [PubMed]  [DOI]  [Cited in This Article: ]
17.  Iakovou I, Schmidt T, Bonizzoni E, Ge L, Sangiorgi GM, Stankovic G, Airoldi F, Chieffo A, Montorfano M, Carlino M. Incidence, predictors, and outcome of thrombosis after successful implantation of drug-eluting stents. JAMA. 2005;293:2126-2130.  [PubMed]  [DOI]  [Cited in This Article: ]
18.  Spertus JA, Kettelkamp R, Vance C, Decker C, Jones PG, Rumsfeld JS, Messenger JC, Khanal S, Peterson ED, Bach RG. Prevalence, predictors, and outcomes of premature discontinuation of thienopyridine therapy after drug-eluting stent placement: results from the PREMIER registry. Circulation. 2006;113:2803-2809.  [PubMed]  [DOI]  [Cited in This Article: ]
19.  Rodriguez AE, Mieres J, Fernandez-Pereira C, Vigo CF, Rodriguez-Alemparte M, Berrocal D, Grinfeld L, Palacios I. Coronary stent thrombosis in the current drug-eluting stent era: insights from the ERACI III trial. J Am Coll Cardiol. 2006;47:205-207.  [PubMed]  [DOI]  [Cited in This Article: ]
20.  Rodriguez AE, Rodriguez-Granillo GA, Palacios IF. Late stent thrombosis: the Damocle's sword of drug eluting stents. EuroIntervention. 2007;2:512-517.  [PubMed]  [DOI]  [Cited in This Article: ]
21.  Daemen J, Wenaweser P, Tsuchida K, Abrecht L, Vaina S, Morger C, Kukreja N, Jüni P, Sianos G, Hellige G. Early and late coronary stent thrombosis of sirolimus-eluting and paclitaxel-eluting stents in routine clinical practice: data from a large two-institutional cohort study. Lancet. 2007;369:667-678.  [PubMed]  [DOI]  [Cited in This Article: ]
22.  Byrne RA, Iijima R, Mehilli J, Pinieck S, Bruskina O, Schömig A, Kastrati A. Durability of antirestenotic efficacy in drug-eluting stents with and without permanent polymer. JACC Cardiovasc Interv. 2009;2:291-299.  [PubMed]  [DOI]  [Cited in This Article: ]
23.  Finn AV, Nakazawa G, Kolodgie FD, Virmani R. Temporal course of neointimal formation after drug-eluting stent placement: is our understanding of restenosis changing. JACC Cardiovasc Interv. 2009;2:300-302.  [PubMed]  [DOI]  [Cited in This Article: ]
24.  Jiménez-Quevedo P, Sabaté M, Angiolillo DJ, Costa MA, Alfonso F, Gómez-Hospital JA, Hernández-Antolín R, Bañuelos C, Goicolea J, Fernández-Avilés F. Vascular effects of sirolimus-eluting versus bare-metal stents in diabetic patients: three-dimensional ultrasound results of the Diabetes and Sirolimus-Eluting Stent (DIABETES) Trial. J Am Coll Cardiol. 2006;47:2172-2179.  [PubMed]  [DOI]  [Cited in This Article: ]
25.  Grube E, Dawkins K, Guagliumi G, Banning A, Zmudka K, Colombo A, Thuesen L, Hauptman K, Marco J, Wijns W. TAXUS VI final 5-year results: a multicentre, randomised trial comparing polymer-based moderate-release paclitaxel-eluting stent with a bare metal stent for treatment of long, complex coronary artery lesions. EuroIntervention. 2009;4:572-577.  [PubMed]  [DOI]  [Cited in This Article: ]
26.  Kukreja N, Onuma Y, Daemen J, Serruys PW. The future of drug-eluting stents. Pharmacol Res. 2008;57:171-180.  [PubMed]  [DOI]  [Cited in This Article: ]
27.  Finn AV, Kolodgie FD, Harnek J, Guerrero LJ, Acampado E, Tefera K, Skorija K, Weber DK, Gold HK, Virmani R. Differential response of delayed healing and persistent inflammation at sites of overlapping sirolimus- or paclitaxel-eluting stents. Circulation. 2005;112:270-278.  [PubMed]  [DOI]  [Cited in This Article: ]
28.  Windecker S, Serruys PW, Wandel S, Buszman P, Trznadel S, Linke A, Lenk K, Ischinger T, Klauss V, Eberli F. Biolimus-eluting stent with biodegradable polymer versus sirolimus-eluting stent with durable polymer for coronary revascularisation (LEADERS): a randomised non-inferiority trial. Lancet. 2008;372:1163-1173.  [PubMed]  [DOI]  [Cited in This Article: ]
29.  Rodriguez AE. Emerging drugs for coronary restenosis: the role of systemic oral agents the in stent era. Expert Opin Emerg Drugs. 2009;14:561-576.  [PubMed]  [DOI]  [Cited in This Article: ]
30.  Aoki J, Ong AT, Abizaid A, den Heijer P, Bonnier H, McClean DR, Verheye S, Belardi J, Condado JA, Pieper M. One-year clinical outcome of various doses and pharmacokinetic release formulations of paclitaxel eluted from an erodable polymer - Insight in the Paclitaxel In-Stent Controlled Elution Study (PISCES). EuroIntervention. 2005;1:165-172.  [PubMed]  [DOI]  [Cited in This Article: ]
31.  Onuma Y, Serruys P, den Heijer P, Joesoef KS, Duckers H, Regar E, Kukreja N, Tanimoto S, Garcia-Garcia HM, van Beusekom H. MAHOROBA, first-in-man study: 6-month results of a biodegradable polymer sustained release tacrolimus-eluting stent in de novo coronary stenoses. Eur Heart J. 2009;30:1477-1485.  [PubMed]  [DOI]  [Cited in This Article: ]
32.  Han Y, Jing Q, Xu B, Yang L, Liu H, Shang X, Jiang T, Li Z, Zhang H, Li H. Safety and efficacy of biodegradable polymer-coated sirolimus-eluting stents in "real-world" practice: 18-month clinical and 9-month angiographic outcomes. JACC Cardiovasc Interv. 2009;2:303-309.  [PubMed]  [DOI]  [Cited in This Article: ]
33.  Ostojic M, Sagic D, Beleslin B, Jung R, Perisic Z, Jagic N, Nedeljkovic M, Mangovski L, Milosavljevic B, Stojkovic S. First clinical comparison of Nobori -Biolimus A9 eluting stents with Cypher- Sirolimus eluting stents: Nobori Core nine months angiographic and one year clinical outcomes. EuroIntervention. 2008;3:574-579.  [PubMed]  [DOI]  [Cited in This Article: ]
34.  Chevalier B, Silber S, Park SJ, Garcia E, Schuler G, Suryapranata H, Koolen J, Hauptmann KE, Wijns W, Morice MC. Randomized comparison of the Nobori Biolimus A9-eluting coronary stent with the Taxus Liberté paclitaxel-eluting coronary stent in patients with stenosis in native coronary arteries: the NOBORI 1 trial--Phase 2. Circ Cardiovasc Interv. 2009;2:188-195.  [PubMed]  [DOI]  [Cited in This Article: ]
35.  Mehilli J, Byrne RA, Wieczorek A, Iijima R, Schulz S, Bruskina O, Pache J, Wessely R, Schömig A, Kastrati A. Randomized trial of three rapamycin-eluting stents with different coating strategies for the reduction of coronary restenosis. Eur Heart J. 2008;29:1975-1982.  [PubMed]  [DOI]  [Cited in This Article: ]
36.  Byrne RA, Kastrati A, Kufner S, Massberg S, Birkmeier KA, Laugwitz KL, Schulz S, Pache J, Fusaro M, Seyfarth M. Randomized, non-inferiority trial of three limus agent-eluting stents with different polymer coatings: the Intracoronary Stenting and Angiographic Results: Test Efficacy of 3 Limus-Eluting Stents (ISAR-TEST-4) Trial. Eur Heart J. 2009;30:2441-2449.  [PubMed]  [DOI]  [Cited in This Article: ]
37.  Byrne RA, Kastrati A. Biodegradable polymer limus-eluting stents are noninferior to permanent polymer-based stents: the ISAR-TEST-4 trial. Interv Cardiol. 2010;2:267-273.  [PubMed]  [DOI]  [Cited in This Article: ]
38.  Lemos PA, Moulin B, Perin MA, Oliveira LA, Arruda JA, Lima VC, Lima AA, Caramori PR, Medeiros CR, Barbosa MR. Randomized evaluation of two drug-eluting stents with identical metallic platform and biodegradable polymer but different agents (paclitaxel or sirolimus) compared against bare stents: 1-year results of the PAINT trial. Catheter Cardiovasc Interv. 2009;74:665-673.  [PubMed]  [DOI]  [Cited in This Article: ]
39.  Rodriguez AE, Vigo CF, Delacasa A, Mieres J, Fernandez-Pereira C, Bernardi V, Bettinoti M, Rodriguez-Granillo AM, Rodriguez-Granillo G, Santaera O. Efficacy and safety of a double-coated paclitaxel-eluting coronary stent: The EUCATAX trial. Catheter Cardiovasc Interv. 2011;77:335-342.  [PubMed]  [DOI]  [Cited in This Article: ]
40.  Ozbek C, Mailänder C, Schilling U, Bach R. Clinical efficacy of a DE stent after an electively planned percutaneous coronary intervention under "real-life" conditions: prospective registry (first-in-man data). Clin Hemorheol Microcirc. 2008;39:311-321.  [PubMed]  [DOI]  [Cited in This Article: ]
41.  Hoffmann M, Horres R, Keller R, Baumann H. Isolation and characterisation of endothelial cell surface heparan sulphate from whole bovine lung for coating of biomaterials to improve haemocompatibility. Biomedical polymers and polymer therapeutics. New York, Boston, Dordrecht, London, Moscow: Kluwer Academic Publishers 2002; 213-226.  [PubMed]  [DOI]  [Cited in This Article: ]
42.  Kastrati A, Dibra A, Spaulding C, Laarman GJ, Menichelli M, Valgimigli M, Di Lorenzo E, Kaiser C, Tierala I, Mehilli J. Meta-analysis of randomized trials on drug-eluting stents vs. bare-metal stents in patients with acute myocardial infarction. Eur Heart J. 2007;28:2706-2713.  [PubMed]  [DOI]  [Cited in This Article: ]
43.  Perez G, Rodriguez-Granillo AM, Mieres J, Llaurado C, Rubilar B, Risau G, Fernandez-Pereira C, Rodriguez AE. New coating stent design for patients with high-risk coronary lesions for thrombotic events: early and long-term results of the Camouflage registry. J Invasive Cardiol. 2009;21:378-382.  [PubMed]  [DOI]  [Cited in This Article: ]