Copyright
©The Author(s) 2024.
World J Gastrointest Endosc. Jan 16, 2024; 16(1): 37-43
Published online Jan 16, 2024. doi: 10.4253/wjge.v16.i1.37
Published online Jan 16, 2024. doi: 10.4253/wjge.v16.i1.37
Table 1 Reported procedure type in which gold probe was used
Procedure | Number |
Bronchoscopy | 7 |
Colonoscopy | 25 |
Esophagogastroduodenoscopy | 47 |
Enteroscopy | 3 |
Endoscopic retrograde cholangiopancreatography | 6 |
Missing | 52 |
Total | 140 |
Table 2 Reported procedure type in which injection gold probe was used
Procedure | Number |
EGD | 174 |
Colonoscopy | 16 |
ERCP | 11 |
Enteroscopy | 1 |
Total | 202 |
Table 3 Failure Modes of gold probe (Primary Outcomes)
Failure mode | Number | Percentage |
Arcing | 1 | 0.7 |
Component missing | 1 | 0.7 |
Detachment of device component | 1 | 0.7 |
Electrical connector broke during use | 1 | 0.7 |
Failure to deliver energy | 107 | 76.4 |
Material separation | 28 | 20.0 |
Tip bent (from packaging) | 1 | 0.7 |
Total | 140 | 100.0 |
Table 4 Failure modes of injection gold probe (Primary Outcomes)
Failure mode | Number | Percentage |
Crack | 9 | 4.5 |
Material separation | 34 | 16.8 |
Device detachment | 27 | 13.4 |
Failure to deliver energy | 115 | 56.9 |
Material puncture/hole | 5 | 2.5 |
Mechanical problem | 12 | 5.9 |
Total | 202 | 100.0 |
Table 5 Events affecting patients with gold probe failure (Secondary Outcomes)
Events | Number |
Bowel burn | 1 |
Hemorrhage/Bleeding | 1 |
Prolonged hospital stay | 1 |
Removal requiring a second procedure | 7 |
Total | 10 |
Table 6 Events affecting patients or equipment with injection gold probe failure (Secondary Outcomes)
Events | Number |
Prolonged hemorrhage | 2 |
The secondary procedure to retrieve the detached probe | 48 |
Probe damaged scope | 3 |
Total | 53 |
Table 7 Reports by year (2013-2023)
Year | MAUDE reports for gold probe | MAUDE reports for injection gold probe |
2013 | 32 | 38 |
2014 | 32 | 36 |
2015 | 17 | 33 |
2016 | 26 | 30 |
2017 | 14 | 23 |
2018 | 3 | 6 |
2019 | 3 | 9 |
2020 | 2 | 8 |
2021 | 4 | 7 |
2022 | 1 | 5 |
2023 | 6 | 7 |
Total | 140 | 202 |