Topic Highlight
Copyright ©2014 Baishideng Publishing Group Inc.
World J Gastroenterol. Nov 14, 2014; 20(42): 15476-15491
Published online Nov 14, 2014. doi: 10.3748/wjg.v20.i42.15476
Table 1 Recent developments on hepatitis C virus detection by electrochemical immunosensors
ElectrodeImmunosensor formatAntigenLabelled systemTechniquePerformance of the immunosensorRef.
3-Mercaptopropionic, acid was assembled on gold electrode to form a self-assembled monolayer-Thionine HCV horseradish peroxidase (HRP) antibody-Cyclic voltammetry and Linear scan voltammetryThe linearity of Hepatitis C in the range of 3.2 to 16 mg/L with the correlation coefficient of 0.995[90]
Glassy carbon electrode modified with a nanocomposite (gold/zirconia/chitosan)Sandwich-typeHCV coreLabel-freeCyclic voltammetry, Electrochemical impedance spectroscopyStability: 98.5% (30 d)[91]
Detection limit: 0.17 ng/mLRSD (different electrodes): 5.1%, 3.9% and 4.8% for 2, 32 and 128 ng/mL HCV core antigen
GMCs-MB nanocompositeSandwich-typeHCV coreHorseradish peroxidase-DNA-coated CMWNTs as a secondary antibody layerCyclic voltammetry, Electrochemical impedance spectroscopyStability: 91.6% (10 d)[92]
Detection limit: 0.01 pg/mL
Regeneration: 8 mol/L urea solution (RSD, 3.1%)
Glassy carbon electrode was modified with an Au-MoO3/Chitosan nanocompositeSandwich-typeHCV non-structural 5A proteinHorseradish peroxidase-labelled antibodyCyclic voltammetry, Chronoamperometry, Electrochemical impedance spectroscopyDetection range: 1.0 ng/mL-50 μg/mL[93]
Silk fibroin nanostructured films on screen-printed carbono electrodeDirect assayHCV non-structural 5A proteinLabel-freeCyclic voltammetrySensitivity: 28 μA/μgDetection range: 0 μg/mL-0.2 μg/mL[94]