Observational Study
Copyright ©2014 Baishideng Publishing Group Inc.
World J Gastroenterol. Oct 21, 2014; 20(39): 14472-14478
Published online Oct 21, 2014. doi: 10.3748/wjg.v20.i39.14472
Table 1 Institutions and patients included
Institutionn
Hospital de Navarra8
Clinica Universidad de Navarra5
Hospital Gregorio Marañon10
Hospital Juan Canalejo10
Hospital Puerta de Hierro9
Hospital del Mar8
Complejo Hospitalario de Toledo9
Hospital La Fe10
Hospital General de Valencia10
Mutua de Tarrassa10
Total89
Table 2 PillCam SB24 specifications
Physical dimensions26 mm × 11 mm
Weight2.89 ± 0.15 g
Number of imaging heads1
Imager typeCMOS
Image capture rate4 frames per second
Optical illumination4 white light emitting diodes with ALC
Field of view156° - working distance of 4.5 mm
Depth of field0-30 mm
Mucosa area images at 4.5 mmApproximately 1100 mm2
Magnification1.8
Min. detectable object size0.1 mm
Min. operation time7 h and 58 min
Max. operation time9:00 h
Shelf life10 mo after manufacturing date
Transmitter frequency434.1 MHz
Battery typeSilver oxide, non-toxic
Storage temperature0°-30°
ActivationMagnetic, automatic from blister
Table 3 Overall number of lesions and abnormal frames detected n (%)
LesionsFrames
TotalTotal
Esophagus17 (6.6)132 (4)
Stomach30 (11.6)617 (19.1)
Small bowel193 (75)2227 (69)
Colon17 (6.6)249 (7.7)
Total257 (100)3225 (100)
Table 4 Comparative analysis of landmark detection n (%)
SB24 patients“SB2-like” patientsP valueK value
Z-line53 (59.5)45 (50.6)NS0.820
Papilla7 (7.9)6 (6.7)NS0.917
Table 5 Comparative analysis of frames per landmark n (%)
SB24
"SB2-like"
P value
FramesMean ± SDFramesMean ± SD
Z-line61611.6 ± 20.43918.6 ± 10.6NS
Papilla355.0 ± 2.0152.5 ± 1.4< 0.05
Table 6 Comparative analysis of quality of images detected
Quality1SB24“SB2-like”P value
Mean ± SDMean ± SD
91.94 ± 21.8288.01 ± 34.87NS
Table 7 Lesions detection: Per-patient analysis n (%)
SB24 patients1“SB2-like” patients1P valueK value
Esophagus15 (23.0)14 (22.2)NS0.9
Stomach18 (27.7)18 (28.5)NS1.0
Small bowel59 (90.7)59 (93.6)NS1.0
Colon11 (16.9)11 (17.4)NS0.9
Total65 (100)64 (100)NS0.9
Table 8 Lesions detection: Per-lesion analysis
SB24
“SB2-like”
P value
LesionsMean ± SDLesionsMean ± SD
Esophagus170.2 ± 0.4140.1 ± 0.4NS
Stomach300.3 ± 0.8300.3 ± 0.8NS
Small bowel1932.2 ± 3.11832.1 ± 2.8NS
Colon170.2 ± 0.6170.2 ± 0.6NS
Total2572.9 ± 3.82442.7 ± 3.5NS
Table 9 Comparative analysis of frames per lesion
SB24
“SB2-like”
P value
FramesMean ± SDFramesMean ± SD
Esophagus1327.7 ± 9.0916.5 ± 8.7NS
Stomach61720.5 ± 34.240313.4 ± 25.8NS
Small Bowel222711.5 ± 28.114908.1 ± 18.1NS
Colon24914.6 ± 18.21559.1 ± 12.4NS
Total322512.8 ± 15.221398.7 ± 15.6NS