Preliminary study of a dot immunogold filtration assay for rapid detection of anti-HCV IgG
Le-Yi Xiao, Xiao-Jun Yan, Feng-Chan Han, Yu Hou, Institute of Genetic Diagnosis of Chinese People′s Liberation Army, Fourth Military Medical University, Xi′an 710033, China
Ming-Ren Mi, Xi′an Railway Central Hospital, Xi′an 710054, China
Author contributions: All authors contributed equally to the work.
Correspondence to: Dr. Le-Yi Xiao, Institute of Genetic Diagnos is of Chinese People′s-Liberation Army, Fourth Military Medical University, Xi′an 710033, China
Telephone: +86-29-3374771
Received: January 20, 1999
Revised: May 3, 1999
Accepted: May 19, 1999
Published online: August 15, 1999

Abstract
Key Words: hepatitis C virus; IgG/analysis; dot immunogold; filtration assay



INTRODUCTION

Hepatitis C is a world-wide epidemic disease stemming from the hepatitis C virus (HCV). HCV is not only the pathogenic factor of hepatitis C but also plays an important role in the process of triggering and development of cirrhosis and liver cancer. There are two indicators for detecting HCV infection: one is HCV RNA which is the gene of the virus itself and the other is anti-HCV which is the antibody produced when the HCV antigen stimulates the body. The HCV antibody detection includes the detection of HCV IgG and HCV-IgM antibody. At present, EIA and the recombinant immunoblot assay (RIBA) are mainly used for detecting anti-HCV-IgG. By using the colloidal gold labeling technique to fix the structural and the non-structural domains of HCV recombinant antigen on the millipore membrane, a dot immunogold filtration assay for detecting anti-HCV-IgG was successfully established.

MATERIALS AND METHODS
Materials

Nitrocellulose (NC) membrane, 0.6 μm diameter-pore, is the product of the Beijing Chemical School. HCV recombinant antigen (rHcAg) is a gift from Li Yuexi of the Nanjing Military Medical Research Institute. Chloroauric acid is the product of the Chengdu Chemical Reagent Factory (batch number 930821). Anti-HCV antibody EIA kit is produced by the Nanjing Military Medical Research Institute. Serum specimens were provided by the Clinical Lab and the Blood-Transfusion Center of Xijing Hospital. Staphylococcus A protein (SPA) is the product of Shanghai Biological Products Research Institute (batch number 941201).

Methods

Colloidal gold preparation. The colloidal gold was prepared by the sodium citrate reduction method[1-4]. Fifty mL of the 0.2 mL/L chloroauric acid was heated to the boiling point, then added 1.2 mL sodium citrate solution under magnetic mixing, kept at boiling point for 5 min until the mixture turned dark red.

SPA labelling. The pH value of 1 mL colloidal gold was adjusted to 5.9-6.2 with 0.1 mol/L K2CO3, and added with 6 μg of SPA, stirred for 10 min, then added with the bovine serum albumin (BSA) to a final concentration of 1%. The precipitate was discarded after low-speed centrifugation, the supernatant was contrifuged at 12000 ×g for 20 min and discarded, the precipitate was redissolved in 0.01 mol/L PBS (pH7.4) and the working solution of the gold labeled SPA was prepared and stored at 4 °C for use.

Dot immunogold filtration assay (DIGFA). One μL rHcAg in 0.5 mol/L carbonate buffer (pH9.5) was dotted on NC membrane, dried at room temperature and diluted coated with 5 mL/L BSA for 30 min and washed 10 min × 2 times with 0.01 mol/L PBS-T. After that, the membrane was put on filter paper to remove the buffer, dried and placed into a self-ma de immune filtration device. Before use, a drop of 0.01 mol/L PBS-T was applied to activate the membrane surface at first. After fluid absorbent, 5 μL serum, 2-3 drops of PBS-T and 30 μL of gold labeled SPA were added sequencially. When the permeation completed, 2-3 drops of PBS-T were added to observe the results, red dots mean positive.

Blocking test Ten positive controls were randomly chosen and equal amount of rHcAg with a fixed concentration was added. The samples were kept at 37 °C water both for 1 h and centrifuged at 800 ×g for 20 s, 10 μL of each sample was used for the DIGFA test.

Detection by EIA It was performed according to the operation instructions.

RESULTS

Comparison of DIGFA and EIA A total of 131 samples were detected at the same time by EIGFA and EIA (Table 1). Taking the results of EIA as standard, the sensitivity of DIGFA was 98% and specificity was 98%. The agreement between the two methods was 97%.

Table 1 Comparison of DIGFA and EIA.
DIGFAEIA
Total
PositiveNegative
Positive49352
Negative17879
Total5081131

Blocking test To confirm the specificity of DIGFA, a blocking test was conducted in 10 positive control sera, and all turned to negative. To further testify the sensitivity of DIGFA, a dilution test was made in 10 samples of positive controls at random. The highest and the lowest titer were 1:256 and 1:32, respectively. Moreover, identical results were obtained when the 10 samples were tested for 3 times by DIGFA.

DISCUSSION

It is quite difficult to directly detect the HCV antigen because the virus concentration in the blood of infected person is extremely low. Therefore, it is of great practical importance to establish a specific, sensitive, simple and convenient method for diagnosing HCV-infected persons, screening blood donors and preventing hepatitis C from spreading.

In 1989, Chiron Company developed the first-generation anti-HCV detection kit by applying the gene engineering to express the HCV-encoded 5-1-1 and C100-3 antigen[5,6]. Soon after the second generation anti-HCV EIA kit was developed to enhance its specificity and sensitivity[7,8]. At present, Ortho Company has produced the third generation EIA kit, the sensitivity and specificity of the anti HCV were further improved. The difference of the three generation kits was in the varied coated antigens. This means that the difference is caused b y the varied antigens[8]. At present, the second generation EIA kit is being widely used in detecting anti-HCV around the world. Although better specificity and sensitivity for detecting anti-HCV by EIA can be obtained, the procedure is comparatively complicated, the process is long, and special instruments a re required, it is rather difficult for an ordinary laboratory to carry out the detection. Since SPA can bind human IgG Fc fragment, we used the structural and non-structural domains of recombinant HCV antigens and established a new method for the rapid detection of HCV antibody based on the mechanism of dot immunogold filtration assay. It showed that the specificity and sensitivity of the method are basically equal to EIA. Furthermore, the new method is simple, convenient and rapid. The required amount of sera is less and no special instrument is needed. Using the same antigen coating as EIA, it is easy to get the reagents because anti-human IgG used in EIA is replaced by SPA. Moreover, the easier preparation of SPA do not damage the activity of IgG so that the labeling efficiency is higher. In dot immunogold filtration technique, the reaction is on membrane, the operation is much simple and convenient and the results can be visually read. As a result, the method can be applied for the serological diagnosis and epidemiological investigation of hepatitis C.

Footnotes

Dr. Le-Yi Xiao, male, born on 1955-08-10 in Tianmen City, Hubei Province, graduated from Fourth Military Medical University, majoring molecular biology and biochemistry, having 23 papers published.

Edited by MA Jing-Yun

References
1.  Dar VS, Ghosh S, Broor S. Rapid detection of rotavirus by using colloidal gold particles labeled with monoclonal antibody. J Virol Methods. 1994;47:51-58.  [PubMed]  [DOI]  [Cited in This Article: ]  [Cited by in Crossref: 11]  [Cited by in F6Publishing: 11]  [Article Influence: 0.4]  [Reference Citation Analysis (0)]
2.  Spielberg F, Kabeya CM, Ryder RW, Kifuani NK, Harris J, Bender TR, Heyward WL, Quinn TC. Field testing and comparative evaluation of rapid, visually read screening assays for antibody to human immunodeficiency virus. Lancet. 1989;1:580-584.  [PubMed]  [DOI]  [Cited in This Article: ]
3.  Xiao LY, Yan XJ, Chen YX, Su CZ, Xu DZ, Li SQ, Yan YP, Guo YH. Primary study of a dot mmunogold filtration assay for rapid detection of HBV anti-HBs, anti-HBe and HBc antibody. J Fourth Milit Med Univ. 1995;16:250.  [PubMed]  [DOI]  [Cited in This Article: ]
4.  Cai WQ, Wang BY.  Practical Immunocytochemistry. First Edition. Chendu: Sichuan Science and Technology Publishing House 1988; 168-180.  [PubMed]  [DOI]  [Cited in This Article: ]
5.  Choo QL, Kuo G, Weiner AJ, Overby LR, Bradley DW, Houghton M. Isolation of a cDNA clone derived from a blood-borne non-A, non-B viral hepatitis genome. Science. 1989;244:359-362.  [PubMed]  [DOI]  [Cited in This Article: ]
6.  Kuo G, Choo QL, Alter HJ, Gitnick GL, Redeker AG, Purcell RH, Miyamura T, Dienstag JL, Alter MJ, Stevens CE. An assay for circulating antibodies to a major etiologic virus of human non-A, non-B hepatitis. Science. 1989;244:362-364.  [PubMed]  [DOI]  [Cited in This Article: ]
7.  Kotwal GJ, Baroudy BM, Kuramoto IK, McDonald FF, Schiff GM, Holland PV, Zeldis JB. Detection of acute hepatitis C virus infection by ELISA using a synthetic peptide comprising a structural epitope. Proc Natl Acad Sci USA. 1992;89:4486-4489.  [PubMed]  [DOI]  [Cited in This Article: ]
8.  Li HM, Wang XT. HCV antibody diagnosis reagent and present situation and prospect of the study of HCV vaccine. China Microbiol Immunol J. 1995;15:128-130.  [PubMed]  [DOI]  [Cited in This Article: ]