Prospective Study
Copyright ©The Author(s) 2024.
Artif Intell Gastrointest Endosc. Mar 8, 2024; 5(1): 90574
Published online Mar 8, 2024. doi: 10.37126/aige.v5.i1.90574
Table 3 Diagnostic performances of artificial intelligence for ColoRectal polyps in different image enhancement modes
AI4CRP (n = 51)

BLI, % (95%CI)
HDWL, % (95%CI)
LCI, % (95%CI)
Multimodal imaging, % (95%CI)
Sensitivity82.1 (0.66-0.92)59.0 (0.42-0.74)76.9 (0.60-0.88)71.8 (0.55-0.84)
Specificity75.0 (0.43-0.93)91.7 (0.60-1.00)83.3 (0.51-0.97)91.7 (0.60-1.00)
PPV91.4 (0.76-0.98)95.8 (0.77-1.00)93.8 (0.78-0.99)96.6 (0.80-1.00)
NPV56.3 (0.31-0.79)40.7 (0.23-0.61)52.6 (0.29-0.75)50.0 (0.29-0.71)
Diagnostic accuracy80.4 (0.66-0.90)66.7 (0.52-0.79)78.4 (0.64-0.88)76.5 (0.62-0.87)

  • Citation: van der Zander QEW, Schreuder RM, Thijssen A, Kusters CHJ, Dehghani N, Scheeve T, Winkens B, van der Ende - van Loon MCM, de With PHN, van der Sommen F, Masclee AAM, Schoon EJ. Artificial intelligence for characterization of diminutive colorectal polyps: A feasibility study comparing two computer-aided diagnosis systems. Artif Intell Gastrointest Endosc 2024; 5(1): 90574
  • URL: https://www.wjgnet.com/2689-7164/full/v5/i1/90574.htm
  • DOI: https://dx.doi.org/10.37126/aige.v5.i1.90574