World J Clin Cases. Sep 6, 2020; 8(17): 3730-3742
Published online Sep 6, 2020. doi: 10.12998/wjcc.v8.i17.3730
Published online Sep 6, 2020. doi: 10.12998/wjcc.v8.i17.3730
Parameters | AUROC | Cutoff Value (kPa) | Sensitivity, % | Specificity, % | PPV, % | NPV, % | LR+ | LR- |
SWE | 0.786 | 9.05 | 77.78 | 77.78 | 77.78 | 77.78 | 3.50 | 0.29 |
TE | 0.714 | 8.15 | 59.26 | 66.67 | 64.00 | 62.07 | 1.78 | 0.61 |
FIB-4 | 0.551 | 0.91 | 74.07 | 44.44 | 57.14 | 63.16 | 1.33 | 0.58 |
APRI | 0.556 | 0.37 | 81.48 | 44.44 | 59.46 | 70.59 | 1.47 | 0.42 |
- Citation: Yao TT, Pan J, Qian JD, Cheng H, Wang Y, Wang GQ. Shear wave elastography may be sensitive and more precise than transient elastography in predicting significant fibrosis. World J Clin Cases 2020; 8(17): 3730-3742