Data acquisition parameter | Minimum requirement | Comments |
Field signal | 1.5 T or 3T | Give a higher visual score, larger number of fibres. 7T and above – enables identification of smaller anatomical structures |
Number of diffusion direction | Approximate 21 | Lowest should be 6 diffusion direction, however it is advice to use more than 6 direction and can go up to 100 directions |
Image resolution | depend on the FOV (usually 24 cm × 24 cm) | Large enough to prevent aliasing |
B-value | 0-1000 s/mm2 | 500-5000+ for HARDI acquisition 2500 for q-space even higher, e.g., 5-8000 |
Number of b-values (b = 0) images | 2 | About 1 per 6 diffusion images |
Number of averages | NSA = 2 | Can improve the SNR |