World J Clin Cases. Aug 6, 2022; 10(22): 7794-7807
Published online Aug 6, 2022. doi: 10.12998/wjcc.v10.i22.7794
Published online Aug 6, 2022. doi: 10.12998/wjcc.v10.i22.7794
Table 3 Patient experience of FreeStyle Libre and Dexcom G5 measured on a visual analogue scale
Variable | FreeStyle Libre isCGM, n = 31 | Dexcom G5 CGM, n = 31 | Change from FreeStyle Libre isCGM to Dexcom G5 CGM | |
P value | ||||
My experience of the system was very positive | 8.35 (2.03) | 6.84 (2.70) | -1.52 (3.41) | 0.023 |
9.00 (3.00; 10.00) | 8.00 (2.00; 10.00) | -2.00 (-7.00; 7.00) | ||
(7.61; 9.10) | (5.85; 7.83) | (-2.77; -0.26) | ||
n = 31 | n = 31 | n = 31 | ||
The insertion of the sensor was easy | 9.03 (1.71) | 8.27 (2.29) | -0.767 (1.455) | 0.0084 |
10.00 (2.00; 10.00) | 9.00 (1.00; 10.00) | 0.000 (-5.000; 2.000) | ||
(8.39; 9.67) | (7.41; 9.12) | (-1.310; -0.223) | ||
n = 30 | n = 30 | n = 30 | ||
I felt safe during my time using the system | 7.94 (2.67) | 7.19 (2.46) | -0.742 (3.916) | 0.32 |
9.00 (0.00; 10.00) | 8.00 (3.00; 10.00) | -1.000 (-7.000; 9.000) | ||
(6.96; 8.91) | (6.29; 8.09) | (-2.178; 0.694) | ||
n = 31 | n = 31 | n = 31 | ||
It was easy to use the system | 9.52 (0.96) | 7.60 (2.62) | -1.93 (2.21) | < 0.0001 |
10.00 (6.00; 10.00) | 9.00 (2.00; 10.00) | -1.00 (-7.00; 0.00) | ||
(9.16; 9.87) | (6.62; 8.58) | (-2.76; -1.11) | ||
n = 31 | n = 30 | n = 30 | ||
It was easy to interpret the information on the receiver screen | 9.42 (0.99) | 7.97 (2.46) | -1.45 (1.98) | < 0.0001 |
10.00 (7.00; 10.00) | 9.00 (1.00; 10.00) | -1.00 (-7.00; 1.00) | ||
(9.06; 9.78) | (7.07; 8.87) | (-2.18; -0.73) | ||
n = 31 | n = 31 | n = 31 | ||
I was not in pain or had discomfort in connection to my use of the system | 9.74 (0.73) | 8.48 (2.78) | -1.26 (2.68) | 0.0078 |
10.00 (7.00; 10.00) | 10.00 (0.00; 10.00) | 0.00 (-10.00; 0.00) | ||
(9.47; 10.01) | (7.46; 9.50) | (-2.24; -0.27) | ||
n = 31 | n = 31 | n = 31 | ||
I experienced no problem scanning/contact with the system | 9.55 (0.93) | 7.19 (3.29) | -2.35 (3.23) | < 0.0001 |
10.00 (7.00; 10.00) | 9.00 (0.00; 10.00) | -1.00 (-10.00; 2.00) | ||
(9.21; 9.89) | (5.99; 8.40) | (-3.54; -1.17) | ||
n = 31 | n = 31 | n = 31 | ||
The sensor was comfortable to have on my body in my daily life | 9.20 (1.40) | 7.23 (2.69) | -1.90 (2.34) | < 0.0001 |
10.00 (4.00; 10.00) | 8.00 (1.00; 10.00) | -1.00 (-8.00; 1.00) | ||
(8.68; 9.72) | (6.24; 8.21) | (-2.77; -1.03) | ||
n = 30 | n = 31 | n = 30 | ||
The system did not disturb my daily life | 9.33 (1.09) | 7.74 (2.62) | -1.50 (2.43) | 0.0024 |
10.00 (6.00; 10.00) | 9.00 (3.00; 10.00) | 0.00 (-7.00; 4.00) | ||
(8.93; 9.74) | (6.78; 8.70) | (-2.41; -0.59) | ||
n = 30 | n = 31 | n = 30 | ||
I would like to use the system in my daily life | 8.45 (2.86) | 5.42 (3.49) | -2.66 (4.98) | 0.0096 |
10.00 (0.00; 10.00) | 6.00 (0.00; 10.00) | -3.00 (-10.00; 9.00) | ||
(7.36; 9.54) | (4.14; 6.70) | (-4.55; -0.76) | ||
n = 29 | n = 31 | n = 29 | ||
It was easy to calibrate the Dexcom G5 | 8.40 (2.21) | |||
10.00 (3.00; 10.00) | ||||
(7.58; 9.22) | ||||
n = 30 | ||||
The alarms did not disturb my daily life | 7.97 (2.74) | |||
9.00 (0.00; 10.00) | ||||
(6.95; 8.99) | ||||
n = 30 |
- Citation: Ólafsdóttir AF, Andelin M, Saeed A, Sofizadeh S, Hamoodi H, Jansson PA, Lind M. Performance of Dexcom G5 and FreeStyle Libre sensors tested simultaneously in people with type 1 or 2 diabetes and advanced chronic kidney disease. World J Clin Cases 2022; 10(22): 7794-7807
- URL: https://www.wjgnet.com/2307-8960/full/v10/i22/7794.htm
- DOI: https://dx.doi.org/10.12998/wjcc.v10.i22.7794