World J Methodol. Jun 26, 2016; 6(2): 133-142
Published online Jun 26, 2016. doi: 10.5662/wjm.v6.i2.133
Published online Jun 26, 2016. doi: 10.5662/wjm.v6.i2.133
Numerical summary from a diabetic EDTA-blood sample 5/2015 | ||||||
Analyte/method group | Mean | SD | CV% | Min | Max | Number |
HbA1c, mmol/mol | ||||||
Abbott Arhitect enzymatic | 53.64 | 2.45 | 4.6 | 48.4 | 32.1 | 6 |
Axis-Shield Afinion | 54.70 | 2.17 | 4.0 | 51.0 | 58.0 | 44 |
Beckman Coulter | 55.2 | 1 | ||||
Hemoque | 51.4 | 3.98 | 7.7 | 46.0 | 55.0 | 5 |
HPLC Bio-Rad D-10 | 55.00 | 1 | ||||
HPLC Bio-Rad Variant | 56.00 | 3.21 | 5.7 | 54.0 | 58.0 | 2 |
HPLC Tosoh | 56.42 | 1.23 | 2.2 | 35.6 | 59.8 | 33 |
Roche Cobas c Tina-Quant | 54.76 | 2.56 | 4.7 | 33.0 | 57.0 | 13 |
Roche Cobas Integra | 55.0 | 1 | ||||
Rocgw Tina Quant | 56.00 | 1 | ||||
Siemens Advia Centaur | 57.93 | 1 | ||||
Siemens DCA 2000+ and Vant | 55.05 | 1.96 | 3.6 | 51.0 | 58.0 | 33 |
Thermo Scientific Konelab | 54.44 | 2.63 | 4.8 | 48.0 | 58.4 | 24 |
All | 55.05 | 2.17 | 3.9 | 32.1 | 59.8 | 165 |
HbA1c | ||||||
Abbott Arhitect enzymatic | 7.20 | 0.00 | 0.0 | 7.2 | 7.2 | 3 |
Axis-Shield Afinion | 7.17 | 0.21 | 2.9 | 6.9 | 7.5 | 44 |
Beckman Coulter | 7.2 | 1 | ||||
Hemoqe | 6.88 | 0.25 | 3.7 | 6.6 | 7.2 | 6 |
HPLC Bio-Rad D-10 | 7.2 | 1 | ||||
HPLC Bio-Rad Variant | 7.5 | 1 | ||||
HPLC Tosoh | 7.34 | 0.13 | 1.8 | 7.0 | 8.8 | 23 |
Roche Cobas c Tina-Quant | 7.29 | 0.12 | 1.6 | 6.7 | 7.4 | 7 |
Siemens Advia Centaur | 7.45 | 5.8 | 7.0 | 1 | ||
Siemens DCA 2000+ and Vant | 7.19 | 0.18 | 2.6 | 6.8 | 7.5 | 31 |
Thermo Scientific Konelab | 7.13 | 0.25 | 3.5 | 6.6 | 7.5 | 26 |
All | 7.20 | 0.21 | 2.9 | 6.6 | 8.8 | 143 |
- Citation: Penttilä I, Penttilä K, Holm P, Laitinen H, Ranta P, Törrönen J, Rauramaa R. Methods, units and quality requirements for the analysis of haemoglobin A1c in diabetes mellitus. World J Methodol 2016; 6(2): 133-142
- URL: https://www.wjgnet.com/2222-0682/full/v6/i2/133.htm
- DOI: https://dx.doi.org/10.5662/wjm.v6.i2.133