Minireviews
Copyright ©The Author(s) 2016.
World J Methodol. Jun 26, 2016; 6(2): 133-142
Published online Jun 26, 2016. doi: 10.5662/wjm.v6.i2.133
Table 2 The methods commonly used for analyzing hemoglobin A1c according to the surveys conducted by Labquality Ltd. in 2015
Numerical summary from a diabetic EDTA-blood sample 5/2015
Analyte/method groupMeanSDCV%MinMaxNumber
HbA1c, mmol/mol
Abbott Arhitect enzymatic53.642.454.648.432.16
Axis-Shield Afinion54.702.174.051.058.044
Beckman Coulter55.21
Hemoque51.43.987.746.055.05
HPLC Bio-Rad D-1055.001
HPLC Bio-Rad Variant56.003.215.754.058.02
HPLC Tosoh56.421.232.235.659.833
Roche Cobas c Tina-Quant54.762.564.733.057.013
Roche Cobas Integra55.01
Rocgw Tina Quant56.001
Siemens Advia Centaur57.931
Siemens DCA 2000+ and Vant55.051.963.651.058.033
Thermo Scientific Konelab54.442.634.848.058.424
All55.052.173.932.159.8165
HbA1c
Abbott Arhitect enzymatic7.200.000.07.27.23
Axis-Shield Afinion7.170.212.96.97.544
Beckman Coulter7.21
Hemoqe6.880.253.76.67.26
HPLC Bio-Rad D-107.21
HPLC Bio-Rad Variant7.51
HPLC Tosoh7.340.131.87.08.823
Roche Cobas c Tina-Quant7.290.121.66.77.47
Siemens Advia Centaur7.455.87.01
Siemens DCA 2000+ and Vant7.190.182.66.87.531
Thermo Scientific Konelab7.130.253.56.67.526
All7.200.212.96.68.8143