Editorial
Copyright ©The Author(s) 2016.
World J Methodol. Mar 26, 2016; 6(1): 20-24
Published online Mar 26, 2016. doi: 10.5662/wjm.v6.i1.20
Figure 1
Figure 1 Specimen cation-exchange high-performance liquid chromatography output from the Bio-Rad Variant II Turbo instrument (Bio-Rad Laboratories, Hercules, United States) using the β-Thal short programme. Label A indicates the total time of analysis (X-axis) is approximately 5 to 6 min; Label B indicates that the total area of analysis should lie between 1 and 3 million; Label C shows the unknown peaks that may occur, especially in the P2 and P3 regions; Label D depicts the preintegration phase (< 1 min) is not reflected in the table above and should be analyzed on the chromatogram; Label E shows where the problem of HbF concentration being calculated as > 100% is present in this case. Please see text for details of resolution of this problem.