Copyright
©The Author(s) 2023.
World J Virol. Sep 25, 2023; 12(4): 233-241
Published online Sep 25, 2023. doi: 10.5501/wjv.v12.i4.233
Published online Sep 25, 2023. doi: 10.5501/wjv.v12.i4.233
Parameter | HBV | HCV | ||
Assay 1 | Assay 2 | Assay 1 | Assay 2 | |
Sample volume | 550 µL plasma | 650 µL plasma | 550 µL plasma | 650 µL plasma |
Turnaround time | Approximately 1.00 h | Approximately 5.00 h | Approximately 1.25 h | Approximately 5.45 h |
LOD in IU/mL | 8.0 | 20 | 8.0 | ≥ 15 |
LLOQ as log10 IU/mL | 0.9 | 1.3 | 0.9 | 1.5 |
Linear range quantification as log10 IU/mL | 1-9 | 1.3-8.2 | 1-9 | 1.5 |
- Citation: Chooramani G, Samal J, Rani N, Singh G, Agarwal R, Bajpai M, Kumar M, Prasad M, Gupta E. Performance evaluation of NeuMoDx 96 system for hepatitis B and C viral load. World J Virol 2023; 12(4): 233-241
- URL: https://www.wjgnet.com/2220-3249/full/v12/i4/233.htm
- DOI: https://dx.doi.org/10.5501/wjv.v12.i4.233