Copyright
©The Author(s) 2025.
World J Transplant. Jun 18, 2025; 15(2): 99287
Published online Jun 18, 2025. doi: 10.5500/wjt.v15.i2.99287
Published online Jun 18, 2025. doi: 10.5500/wjt.v15.i2.99287
Table 1 Trend of the hemodynamic parameters and lactate measured at beginning, after 2 hours and at the end of the treatment of livers treated with hypothermic machine perfusion connected to cytokines filter vs livers perfused with a standard cycle of hypothermic machine perfusion vs livers stored in iced box
Parameters | Groups | T0 | Tend | P value (Tend vs T0) |
Hepatic artery resistance (mmHg/mL/minute) (mean ± SD) | Sorb | 0.62 ± 0.17 | 0.48 ± 0.07 | 0.11 |
NoSorb | 0.76 ± 0.25 | 0.56 ± 0.01 | 0.18 | |
Tend P value | 1 | |||
Hepatic artery flow (mL/minute) (mean ± SD) | Sorb | 66 ± 33 | 85 ± 9 | 0.11 |
NoSorb | 56 ± 41 | 75 ± 28 | 0.18 | |
Tend P value | 0.4 | |||
Portal vein flow (mL/minute) (mean ± SD) | Sorb | 345 ± 78 | 359 ± 97 | 0.28 |
NoSorb | 321 ± 89 | 317 ± 146 | 0.65 | |
Tend P value | 1 | |||
Lactate (mg/mL) (mean ± SD) | Sorb | 8.83 ± 2.11 | 6.4 ± 0.88 | 0.11 |
NoSorb | 15 ± 0.00 | 5.45 ± 2.05 | 0.18 | |
NoTreat | 15 | |||
Tend P value | 0.13 |
- Citation: Scalera I, Franzin R, Stasi A, Castellaneta A, Fischetti E, Morelli G, Raele M, Panetta E, Kurevija A, Pulga W, Atti M, Gesualdo L. Haemoadsorption cartridge connected to the machine perfusion for donation after circulatory death porcine liver marginal grafts. World J Transplant 2025; 15(2): 99287
- URL: https://www.wjgnet.com/2220-3230/full/v15/i2/99287.htm
- DOI: https://dx.doi.org/10.5500/wjt.v15.i2.99287