Trial | 1 | 2 | 3 | 4 | 5 | 6 | 7 | Total |
Number of BGAs | 72 | 34 | 75 | 73 | 66 | 40 | 74 | 434 |
Detected Manipulations | 77 | 36 | 74 | 73 | 78 | 57 | 109 | 514 |
Detected BGAs (hits) | 69 | 33 | 73 | 70 | 66 | 38 | 73 | 422 |
Missed BGAs (misses) | 3 | 1 | 2 | 3 | 0 | 2 | 1 | 12 |
Other events (other) | 5 | 2 | 1 | 1 | 8 | 15 | 32 | 64 |
False detections (errors) | 3 | 1 | 0 | 2 | 4 | 4 | 4 | 18 |
Sensitivity | 0.96 | 0.97 | 0.97 | 0.96 | 1.00 | 0.95 | 0.99 | 0.97 |
Precision | 0.96 | 0.97 | 1.00 | 0.97 | 0.94 | 0.91 | 0.95 | 0.96 |
F1 Score | 0.96 | 0.97 | 0.99 | 0.97 | 0.97 | 0.93 | 0.97 | 0.97 |
- Citation: Peter J, Klingert W, Klingert K, Thiel K, Wulff D, Königsrainer A, Rosenstiel W, Schenk M. Algorithm-based arterial blood sampling recognition increasing safety in point-of-care diagnostics. World J Crit Care Med 2017; 6(3): 172-178