World J Anesthesiol. Mar 27, 2014; 3(1): 119-123
Published online Mar 27, 2014. doi: 10.5313/wja.v3.i1.119
Published online Mar 27, 2014. doi: 10.5313/wja.v3.i1.119
Table 2 Mean output, standard deviation and standard error of the infusions pumps over a 10 h time period
Pump | n | Mean | StdDev | Std Err | P value1 |
On-Q | 10 | 17.3 | 0.62 | 0.19 | < 0.0001 |
Baxter | 10 | 10.1 | 0.17 | 0.05 | < 0.0001 |
AMBU | 10 | 9.5 | 0.09 | 0.03 | < 0.0001 |
- Citation: LeRiger M, Bhalla T, Martin D, Bettesworth J, Tobias JD. Comparison of flow rate accuracy and consistency between the on-Q, baxter, and ambu pain infusion devices. World J Anesthesiol 2014; 3(1): 119-123
- URL: https://www.wjgnet.com/2218-6182/full/v3/i1/119.htm
- DOI: https://dx.doi.org/10.5313/wja.v3.i1.119