ISO 5833:2002 | ||||||
Bending strength (MPa) | Flexural modulus (MPa) | Compressive strength (MPa) | ||||
Limit | > 50 | > 1800 | > 70 | |||
DePuy CMW® 2G | 75.39 | 67.3 (-8.09) | 3002.75 | 2775 (-227.75) | 91.53 | 80.94 (-10.59) |
81.8 (+6.41) | 3159 (+156.25) | 99.11 (+7.58) | ||||
Palacos® R + G | 65.79 | 61.7 (-4.09) | 2552 | 2383 (-169) | 87.46 | 81.92 (-5.54) |
68.7 (+2.91) | 2659 (+107) | 93.13 (+5.67) | ||||
Palacos® fast R + G | 72.17 | 69.3 (-2.87) | 2995.18 | 2784 (-211.18) | 106.25 | 100.26 (-5.99) |
76.2 (+4.03) | 3118 (+122.82) | 110.51 (+4.26) |
- Citation: Caraan NA, Windhager R, Webb J, Zentgraf N, Kuehn KD. Role of fast-setting cements in arthroplasty: A comparative analysis of characteristics. World J Orthop 2017; 8(12): 881-890
- URL: https://www.wjgnet.com/2218-5836/full/v8/i12/881.htm
- DOI: https://dx.doi.org/10.5312/wjo.v8.i12.881