Copyright
©The Author(s) 2015.
World J Orthop. Oct 18, 2015; 6(9): 705-711
Published online Oct 18, 2015. doi: 10.5312/wjo.v6.i9.705
Published online Oct 18, 2015. doi: 10.5312/wjo.v6.i9.705
Composite bones | Human cadaver bones | |||||||||
1 | 2 | 3 | 4 | 5 | 1 | 2 | 3 | 4 | 5 | |
Load to failure criteria (nm) | 11.91 | 9.40 | 10.97 | 10.67 | 9.64 | 15.10 | 12.59 | 13.05 | 13.02 | 15.38 |
Displacement (mm) | 1.76 | 2.00 | 1.85 | 2.00 | 1.70 | 2.00 | 2.00 | 1.71 | 2.00 | 2.00 |
Stiffness (nm/mm) | 6.80 | 4.69 | 5.95 | 5.33 | 5.69 | 7.55 | 6.31 | 7.62 | 6.52 | 7.72 |
Bone density (g/cm2) | - | - | - | - | - | 0.723 | 0.456 | 0.658 | 0.665 | 0.883 |
Failure mode | Distal fracture | Fracture gap | Distal fracture | Fracture gap | Distal fracture | Fracture gap | Fracture gap | Triceps failure | Fracture gap | Fracture gap |
- Citation: Hamilton Jr DA, Reilly D, Wipf F, Kamineni S. Comminuted olecranon fracture fixation with pre-contoured plate: Comparison of composite and cadaver bones. World J Orthop 2015; 6(9): 705-711
- URL: https://www.wjgnet.com/2218-5836/full/v6/i9/705.htm
- DOI: https://dx.doi.org/10.5312/wjo.v6.i9.705