Randomized Controlled Trial
Copyright ©The Author(s) 2020.
World J Orthop. Oct 18, 2020; 11(10): 442-452
Published online Oct 18, 2020. doi: 10.5312/wjo.v11.i10.442
Table 4 Mean wear rates per year, presented as mean with ranges
Conventional PE (Polyethylene)P valueREXPOLP value
Medial (mm/yr)-0.021 (-0.028 to -0.015)< 0.0010.004 (-0.002 to 0.009)0.21
Proximal (mm/yr)0.033 (0.018 to 0.047)< 0.0010.003 (-0.011 to 0.017)0.66
Volume (mm3/yr)15.94 (2.729 to 29.15)0.02-5.545 (-20.36 to 9.269)0.46
Corrected volume (mm3/yr)18.53 (7.188 to 29.86)0.002-2.142 (-14.13 to 9.841)0.72