Table 2 DNA concentration (ng/µL) using four commercially kits for DNA isolation
PureLink Genomic DNA mini kit (Invitrogen–manual) | iPrep genomic DNA kit (Invitrogen–automated) | MagnaPure LC DNA Isolation kit (Roche–automated) | High Pure FFPE DNA isolation kit (Roche–manual) | |
Mean | 157.24 | 8.72 | 20.39 | 86.64 |
Standard deviation | 62.99 | 6.40 | 21.19 | 43.84 |
Minimum | 37.60 | 0.70 | 0.30 | 4.20 |
Maximum | 316 | 29.80 | 121 | 168.90 |
- Citation: Kovacs Z, Jung I, Csernak E, Szentirmay Z, Banias L, Rigmanyi G, Gurzu S. DNA extraction from paraffin embedded colorectal carcinoma samples: A comparison study of manual vs automated methods, using four commercially kits. World J Clin Oncol 2019; 10(9): 307-317
- URL: https://www.wjgnet.com/2218-4333/full/v10/i9/307.htm
- DOI: https://dx.doi.org/10.5306/wjco.v10.i9.307