Table 7 IMRT/VMAT/tomotherapy three-dimensional dosimetry
Ref. | Dosimeter | Read-out method | Treatment site | Delivery type | Photon energy/number of fields | Comparison on the plane or in volume | Gamma index criteria % dose difference (global/local)/DTA (mm)/threshold (%) | Gamma index passing rate |
[86] | MAGIC | MRI/T2 | Model | IMRT/GKSRS | 6MV/Co-60 | Volume | 3/3/ | 50.3% |
[87] | FX gel | OCT/Vista | Head and Neck | IMRT | 6MV | Volume | 3/3/none | 84.1% |
[88] | BANG3 | MRI/T2 | Prostate | Tomotherapy | 6MV/Arc | Volume | 3/3 | 53% |
[89] | PRESAGE | DLOS | Brain | IMRT | 6MV | Volume | 3/3 | 95% |
[90] | MAGIC-f | MRI/T2 | Prostate | Tomotherapy | 6MV/Arc | Plane | 3/3 | 88.4% |
[91] | nPAG | OCTOPUS-IQ | Prostate | IMRT | /7fields | Volume | 95.3% | |
[10] | PAGAT | MRI/T2 | Pituitary | IMRT | 6MV/7fields | Volume | 2/2 | 99.4% |
[33] | BANG3 | MRI/T2 | Prostate | VMAT (Elekta) | 6MV/Arc | Volume | 3 (global)/3/25 | 95.7% |
[92] | BANG3 | MRI/T2 | Prostate | VMAT (Varian) | 6MV/Arc | Volume | 3 (global)/3/50 | 90.0% |
[93] | PRESAGE | DMOS | Brain | IMRT | 6MV/5fields | Volume | 3 (global)/3/10 | 99.4% |
[94] | NIPAM | MRI/T2 | Eye | IMRT | 6MV/5fields | Plane | 3/3/ | 98.5% |
- Citation: Watanabe Y, Warmington L, Gopishankar N. Three-dimensional radiation dosimetry using polymer gel and solid radiochromic polymer: From basics to clinical applications. World J Radiol 2017; 9(3): 112-125
- URL: https://www.wjgnet.com/1949-8470/full/v9/i3/112.htm
- DOI: https://dx.doi.org/10.4329/wjr.v9.i3.112