Table 7 IMRT/VMAT/tomotherapy three-dimensional dosimetry
Ref.DosimeterRead-out methodTreatment siteDelivery typePhoton energy/number of fieldsComparison on the plane or in volumeGamma index criteria % dose difference (global/local)/DTA (mm)/threshold (%)Gamma index passing rate
[86]MAGICMRI/T2ModelIMRT/GKSRS6MV/Co-60Volume3/3/50.3%
[87]FX gelOCT/VistaHead and NeckIMRT6MVVolume3/3/none84.1%
[88]BANG3MRI/T2ProstateTomotherapy6MV/ArcVolume3/353%
[89]PRESAGEDLOSBrainIMRT6MVVolume3/395%
[90]MAGIC-fMRI/T2ProstateTomotherapy6MV/ArcPlane3/388.4%
[91]nPAGOCTOPUS-IQProstateIMRT/7fieldsVolume95.3%
[10]PAGATMRI/T2PituitaryIMRT6MV/7fieldsVolume2/299.4%
[33]BANG3MRI/T2ProstateVMAT (Elekta)6MV/ArcVolume3 (global)/3/2595.7%
[92]BANG3MRI/T2ProstateVMAT (Varian)6MV/ArcVolume3 (global)/3/5090.0%
[93]PRESAGEDMOSBrainIMRT6MV/5fieldsVolume3 (global)/3/1099.4%
[94]NIPAMMRI/T2EyeIMRT6MV/5fieldsPlane3/3/98.5%