Copyright
©The Author(s) 2016.
World J Radiol. Mar 28, 2016; 8(3): 322-330
Published online Mar 28, 2016. doi: 10.4329/wjr.v8.i3.322
Published online Mar 28, 2016. doi: 10.4329/wjr.v8.i3.322
Pulse sequence | Plane | TR (ms) | TE (ms) | Thk/sp (mm) | Flip angle (°) | BW (Hz/Px) | FOV (cm) | Matrix | TA |
T1 Dixon VIBE AC | Coronal | 3.6 | 1.2 | 3/0 | 10 | 965 | 40-50 | 192 × 121 | 0:19 |
T1 VIBE | Axial | 3.9 | 2.5 | 3/0 | 9 | 720 | 35-45 | 288 × 288 | 0:24 |
T2 HASTE FS | Axial | 1600 | 95 | 5/0 | 9 | 710 | 30-40 | 320 × 260 | 1:08 |
EPI DWI | Axial | 11000 | 70 | 6/0 | N/A | 2084 | 30-40 | 160 × 120 | 3:51 |
STIR | Coronal | 4000 | 48 | 4/0 | 120 | 200 | 35-40 | 320 × 192 | 4:54 |
- Citation: Pugmire BS, Guimaraes AR, Lim R, Friedmann AM, Huang M, Ebb D, Weinstein H, Catalano OA, Mahmood U, Catana C, Gee MS. Simultaneous whole body 18F-fluorodeoxyglucose positron emission tomography magnetic resonance imaging for evaluation of pediatric cancer: Preliminary experience and comparison with 18F-fluorodeoxyglucose positron emission tomography computed tomography. World J Radiol 2016; 8(3): 322-330
- URL: https://www.wjgnet.com/1949-8470/full/v8/i3/322.htm
- DOI: https://dx.doi.org/10.4329/wjr.v8.i3.322