Basic Study
Copyright ©The Author(s) 2016.
World J Radiol. Mar 28, 2016; 8(3): 298-307
Published online Mar 28, 2016. doi: 10.4329/wjr.v8.i3.298
Table 1 Multiple magnetic resonance imaging sequences used for temperature monitoring, characterization and quantification of ablated kidneys
Pulse sequenceTR (ms)TE (ms)ETLrBW (kHz)Flip angle (degree)Slice thicknessMatrix sizeNEXAcquisition time (s)
T2-W thermometry21018.31215.63354256 × 1921100
2D T2-W FSE (pre and post ablation)7300681215.63114256 × 1923400
Perfusion (post ablation)3.71.5NA62.59596 × 961120
2D T1-W FSE (pre and post ablation1)9857.2631.251205192 × 192116
3D LAVA (pre and post ablation1)3.91.5NA62.5155192 × 1923240