Table 8 Estimates of apparent diffusion coefficient values reported in the normal fibroglandular breast tissue using a mono-exponential fit
Ref.SystemField strength (T)SequenceTR/TE (ms/ms)No. of patients/mean age (SD)b-values (s/mm2)Mean ADC (SD) (× 10-3 mm2/s)
Englander et al[181], 1997GE Signa1.5SE(2 or 3)HB/1004/32 (2.6)12.48, 611.31, 199.61, 449.12, 77.97, 311.89, 112.281.64 (0.19)
Partridge et al[17], 2001GE Signa1.5SS FSE8000/868/270, 578n = 6, W1: 1.72 (0.23) W2: 1.61 (0.22) W3: 1.66 (0.17) W4: 1.75 (0.17)
Sinha et al[105], 2002GE Signa1.5SS EPI5000/72160, 269, 537, 806, 10742.37 (0.27)
Woodhamset al[106], 2005GE Signa1.5SS SE EPI5000/61.8190/530, 750, 10002.09 (0.27)
Park et al[182], 2007GE Signa1.5SS EPI6000/7541/53.10, 10001.51 (0.29)
Yoshikawa et al[107], 2008Philips Intera1.5SS SE EPI6238/90270, 200, 400, 600, 8001.96 (0.21)
Kim et al[119], 2009GE Signa1.5SS SE EPI6000/7567/540, 10001.59 (0.27)
Baron et al[114], 2010Siemens Avanto1.5SS SE EPI, (1) FS (2) STIR (3) SPAIR (4) WaterExcitation3000/937/26.60, 20, 40, 60, 80, 100, 150, 200, 300, 400, 500, 600(1) 1.99 (2) 2.02 (3) 2.03 (4) 2.06 overall 2.03 (0.03)
Partridge et al[44], 2010GE Signa1.5SS SE EPI7 s/71.512/36 (median)0, 600, 10001.95 (0.24)
O’Flynn et al[109], 2012Philips Achieva3.0SS SE EPI, SPAIR, SSGR3771/6731/45.9 13 PRM/36.6 18 PM/55.2(1) 0, 100, 150, 200, 350, 700, 1200 (2) 100, 150, 200, 350, 700, 1200PRM, (1) 1.84 (0.26) (2) 1.77 (0.26) PM, (1) 1.46 (0.3) (2) 1.33 (0.3)
Tagliafico et al[183], 2012GE Signa3.0SS EPI16675/NR60/57 (median)0, 10001.92 (0.30)
AlRashidi et al[108], 2012Philips Achieva3.0SS SE EPI, SPAIR9543/50260, 50, 150, 800MC, W1: 2.18 (0.38) W2: 2.20 (0.39) W3: 2.22 (0.38) W4: 2.22 (0.33)
McDonald et al[184], 2014Philips Achieva3.0SS SE EPI5336/61103/47 (11)0, 8001.62 (0.30)