Diagnosis Advances
Copyright ©The Author(s) 2016.
World J Radiol. Jan 28, 2016; 8(1): 1-20
Published online Jan 28, 2016. doi: 10.4329/wjr.v8.i1.1
Table 1 Optimized conventional imaging protocol at 1.5 Tesla (Magnetom Avanto B13, Siemens, Erlangen, Germany), configured with a 12 element designed head matrix coil
1.5 TPlaneSeq.TR (msec)TE/TI (msec)Flip angleSlice (mm)FOVNEXMatrix sizeScan time
NeonatesAxial1SE T1w450-650129032003192 × 2564.30
AxialIR840060/35018032302176 × 5124.30
Coronal AxialTSE T2w89009914532002176 × 2565.00
Axial2GRE T21w640183552002160 × 2563.30
Axial DWI (b0 and b800)EPI5700139-5200496 × 1281.00
Infants and adolescentsAxialFLAIR T2w7000110/200018032003154 × 2564.30
Axial1SE T1w450-650129032003192 × 2564.30
AxialIR840060/35018032302176 × 5124.30
CoronalTSE T2w89009914532002176 × 2565.00
Axial2GRE T21w640183552002160 × 2563.30
Axial DWI (b0 and b1000)EPI5700139-5200496 × 1281.00