Original Article
Copyright ©2013 Baishideng Publishing Group Co.
World J Radiol. Sep 28, 2013; 5(9): 334-344
Published online Sep 28, 2013. doi: 10.4329/wjr.v5.i9.334
Table 2 Imaging parameters in the three imaging systems used for the volunteer population
1.5T MRnon-dpTX 3.0TMRdpTX 3.0TMR
TR/TE, ms6300/796600/806000/68
Sequence typeEPI-SE
FOV, mm × mm380 × 297
Matrix192 × 150
Slice thickness, mm6
Interslice gap, mm0
Spatial resolution, mm³2.0 × 2.0 × 6.0
Number slices35
b-values0, 50, 100, 200, 400, 800
Parallel imagingGRAPPA 2
Acquisition time, min7:027:226:54
Respiratory controlFree breathing
Fat suppressionSPAIR
Averages4
Bandwidth, Hz/px1628