Copyright
©2012 Baishideng Publishing Group Co.
World J Radiol. Jul 28, 2012; 4(7): 328-334
Published online Jul 28, 2012. doi: 10.4329/wjr.v4.i7.328
Published online Jul 28, 2012. doi: 10.4329/wjr.v4.i7.328
Device | Sensitivity | Specificity | PPV | NPV | |
Pre offset | A | 35.48 | 99.41 | 91.67 | 89.42 |
B | 41.94 | 99.41 | 92.86 | 90.37 | |
Post offset | A | 80.65 | 98.26 | 89.29 | 95.57 |
B | 70.97 | 94.12 | 68.75 | 94.67 | |
Mean (A + B) | 77.42 | 94.71 | 72.37 | 95.83 |
- Citation: Haneder S, Gutfleisch A, Meier C, Brade J, Hannak D, Schoenberg SO, Becker CR, Michaely HJ. Evaluation of a handheld creatinine measurement device for real-time determination of serum creatinine in radiology departments. World J Radiol 2012; 4(7): 328-334
- URL: https://www.wjgnet.com/1949-8470/full/v4/i7/328.htm
- DOI: https://dx.doi.org/10.4329/wjr.v4.i7.328