Copyright
©2012 Baishideng Publishing Group Co.
World J Radiol. Jan 28, 2012; 4(1): 21-28
Published online Jan 28, 2012. doi: 10.4329/wjr.v4.i1.21
Published online Jan 28, 2012. doi: 10.4329/wjr.v4.i1.21
Table 3 Weight and maximum transverse diameter distribution of number of subjects with acceptable and unacceptable subjective image noise in abdominal computed tomography examinations for unprocessed and non-linear adaptive filters processed image series at 150, 100 and 50 mAs tube current-time product levels (mean ± SD)
Unprocessed images | Processed images | |||
Acceptable noise | Unacceptable noise | Acceptable noise | Unacceptable noise | |
150 mAs | 14/19 | 5/19 | 19/19 | 0/19 |
Weight (kg) | 74.5 ± 14.8 | 95.2 ± 11.1 | 79.9 ± 16.5 | - |
Transverse diameter (cm) | 34.2 ± 4.3 | 42.9 ± 4.5 | 36.5 ± 5.8 | - |
100 mAs | 4/19 | 15/19 | 17/19 | 2/19 |
Weight (kg) | 77.1 ± 4.0 | 80.7 ± 18.6 | 76.9 ± 14.6 | 106.0 ± 4.3 |
Transverse diameter (cm) | 32.5 ± 3.5 | 37.6 ± 5.9 | 35.7 ± 5.6 | 43.5 ± 0.7 |
50 mAs | 0/19 | 19/19 | 5/19 | 14/19 |
Weight (kg) | - | 79.9 ± 16.5 | 66.2 ± 13.0 | 91.4 ± 15.1 |
Transverse diameter (cm) | - | 36.5 ± 5.8 | 33.0 ± 3.4 | 40.6 ± 6.1 |
- Citation: Singh S, Kalra MK, Sung MK, Back A, Blake MA. Radiation dose reduction with application of non-linear adaptive filters for abdominal CT. World J Radiol 2012; 4(1): 21-28
- URL: https://www.wjgnet.com/1949-8470/full/v4/i1/21.htm
- DOI: https://dx.doi.org/10.4329/wjr.v4.i1.21