Table 4 Correlation of the heavy wave velocities of the Toshiba Aplio 500 version 5.0 and version 6.0, GE Logiq E9, and Philips Epiq 7 devices with the Siemens S3000
Siemens S3000 | |||||
Study arm A (n = 200) | Study arm B (n = 131) | ||||
Device | Pole | R value | P value | R value | P value |
Toshiba Aplio 500 | Lower pole | -0.08792 | 0.2193 | 0.19863 | 0.0375 |
Middle pole | 0.13632 | 0.0561 | 0.13438 | 0.1616 | |
Upper pole | 0.03951 | 0.5815 | 0.24951 | 0.0086 | |
GE Logiq E9 | Lower pole | -0.03307 | 0.6446 | - | - |
Middle pole | -0.04941 | 0.4905 | - | - | |
Upper pole | 0.04744 | 0.5079 | - | - | |
Philips Epiq 7 | Lower pole | 0.04894 | 0.4947 | - | - |
Middle pole | 0.12321 | 0.0845 | - | - | |
Upper pole | 0.33088 | < 0.0001 | - | - |
- Citation: Nowotny F, Schmidberger J, Schlingeloff P, Binzberger A, Kratzer W. Comparison of point and two-dimensional shear wave elastography of the spleen in healthy subjects. World J Radiol 2021; 13(5): 137-148