Guidewire | Core material and coating | Wire guide diameter inch | Wire guide length (cm) | Taper length (cm) | Floppy tip length (cm) | Stiffness | Preshaped curve | Use |
Amplatz Extra Stiff wire (Cook Medical Inc.) | PTFE-coated stainless steel | 0.035 | 260 | 7 | 3 | Least stiff | 3 cm | TAVR device delivery Straightening of tortuous vessels Sheath insertion |
Amplatz Super Stiff wire (Boston Scientific) | PTFE-Coated Stainless Steel | 0.035 | 260 | 6 | 3 | Stiffer than Extra stiff | 1 and 3 cm | Straightening of tortuous vessels Sheath insertion |
Lunderquist Extra Stiff wire (Cook Medical Inc.) | PTFE-Coated Stainless Steel | 0.035 | 260 and 300 | 7.5 | 4 | Stiffer than Amplatz extrastiff and superstiff | 4 cm | Straightening of tortuous vessels Sheath insertion |
Safari Guide wire (Boston Scientific) | LUBRIGREENTM PTFE-Coated Stainless Steel | 0.035 | 260 and 300 | Stiffness equal to Amplatz extrastiff | Small curve - 16 cm distal grind and 1.7 inch/4.25 cm curve Large curve - 18 cm distal grind and 1.9 inch/4.90 cm curve | TAVR device delivery |
- Citation: Chaudhry MA, Sardar MR. Vascular complications of transcatheter aortic valve replacement: A concise literature review. World J Cardiol 2017; 9(7): 574-582
- URL: https://www.wjgnet.com/1949-8462/full/v9/i7/574.htm
- DOI: https://dx.doi.org/10.4330/wjc.v9.i7.574