Table 2 Procedural characteristics
Procedural characteristic | S3-THV (n = 283) | XT-THV (n = 507) | P value |
Transfemoral approach | 232 (82.6) | 273 (53.8) | < 0.0001 |
Local anesthesia | 232 (82.6) | 271 (54.2) | < 0.0001 |
Predilatation | 50 (17.7) | 440 (86.8) | < 0.0001 |
Postdilatation | 45 (15.9) | 61 (12.0) | 0.13 |
Implanted device size | < 0.0001 | ||
23 mm | 111 (39.8) | 127 (25.1) | |
26 mm | 101 (36.2) | 270 (53.4) | |
29 mm | 67 (24.0) | 109 (21.5) | |
Valve area oversizing, % | 11.5 ± 9.8 | 22.9 ± 11.2 | < 0.0001 |
Device diameter/annulus diameter (area-derived) | 1.05 ± 0.05 | 1.11 ± 0.05 | < 0.0001 |
Need for seconde valve implantation | 7 (2.5) | 8 (1.6) | 0.42 |
Annulus rupture | 0 (0) | 13 (2.6) | 0.01 |
Conversion to SAVR | 2 (0.7) | 14 (2.8) | 0.06 |
Contrast use (mL) | 108.2 ± 42.7 | 131.6 ± 60.9 | < 0.0001 |
Fluoroscopy time (min) | 17.4 ± 9.9 | 16.5 ± 9.8 | 0.28 |
- Citation: Sawaya FJ, Spaziano M, Lefèvre T, Roy A, Garot P, Hovasse T, Neylon A, Benamer H, Romano M, Unterseeh T, Morice MC, Chevalier B. Comparison between the SAPIEN S3 and the SAPIEN XT transcatheter heart valves: A single-center experience. World J Cardiol 2016; 8(12): 735-745