Manufacturer | Stent type | Stent size (mm) | Balloon1 (bench test) (mm) | Maximum expansion (bench test) (mm)[72] | Maximum expansion (mm) (manufacturer recommendation) |
Biosensors | Biomatrix A | 3 | 5 | 4.1 | - |
4 | 6 | 5.9 | 5.8 | ||
Medtronic | Resolute Onyx | 2.5 | 4 | 3.3 | - |
3 | 5 | 4.4 | - | ||
4 | 6 | 5.6 | - | ||
5 | 6 | 6 | 6 | ||
Biotronik | Orsiro | 3 | 5 | 4 | - |
4 | 6 | 5.3 | 4.4 | ||
Abbott | Xience | 3 | 5 | 4.1 | 3.75 |
4 | 6 | 5.6 | 5.5 | ||
Terumo | Ultimaster | 3 | 5 | 4.3 | - |
4 | 6 | 5.8 | 5.5 | ||
Boston Scientific | Synergy | 2.75 | 5 | 3.6 | 3.5 |
3.5 | 5 | 4.2 | 4.25 | ||
4 | 6 | 5.7 | 5.75 |