Copyright
©The Author(s) 2024.
World J Gastrointest Surg. Jul 27, 2024; 16(7): 2031-2039
Published online Jul 27, 2024. doi: 10.4240/wjgs.v16.i7.2031
Published online Jul 27, 2024. doi: 10.4240/wjgs.v16.i7.2031
Machine A vs machine B | Machine A vs machine C | Machine B vs machine C | ||||
ICC (95%CI) | P value | ICC (95%CI) | P value | ICC (95%CI) | P value | |
D in PaH | 0.274 (-0.731, 0.697) | 0.232 | 0.342 (-0.411, 0.712) | 0.145 | 0.728 (0.358, 0.886) | 0.002 |
D* in PaH | -0.186 (-1.545, 0.482) | 0.670 | 0.27 (-0.611, 0.685) | 0.220 | 0.578 (0.02, 0.822) | 0.022 |
f in PaH | -0.209 (-1.818, 0.491) | 0.673 | 0.173 (-0.386, 0.582) | 0.272 | 0.19 (-0.541, 0.622) | 0.274 |
D in PaB | 0.456 (-0.332, 0.776) | 0.089 | 0.683 (-0.026, 0.886) | 0.028 | 0.551 (-0.079, 0.814) | 0.037 |
D* in PaB | -0.521 (-3.056, 0.393) | 0.815 | -0.023 (-1.636, 0.587) | 0.519 | -0.256 (-2.207, 0.491) | 0.691 |
f in PaB | 0.385 (-0.407, 0.739) | 0.124 | -0.087 (-1.441, 0.534) | 0.580 | -0.332 (-2.715, 0.515) | 0.748 |
D in PaT | 0.324 (-0.614, 0.718) | 0.186 | 0.286 (-0.557, 0.69) | 0.201 | 0.908 (0.781, 0.961) | < 0.001 |
D* in PaT | 0.569 (-0.204, 0.836) | 0.060 | -0.289 (-2.251, 0.475) | 0.713 | -0.062 (-1.02, 0.504) | 0.569 |
f in PaT | 0.531 (-0.144, 0.806) | 0.047 | 0.35 (-0.633, 0.735) | 0.174 | 0.486 (-0.275, 0.789) | 0.073 |
SNR | 0.081 (-0.206, 0.421) | 0.312 | 0.178 (-0.238, 0.551) | 0.205 | 0.067 (-0.152, 0.364) | 0.301 |
ADC value | 0.870 (0.691, 0.946) | < 0.001 | 0.707 (0.393, 0.871) | < 0.001 | 0.808 (0.590, 0.917) | < 0.001 |
- Citation: Liu X, Wang YF, Qi XH, Zhang ZL, Pan JY, Fan XL, Du Y, Zhai YM, Wang Q. Reproducibility study of intravoxel incoherent motion and apparent diffusion coefficient parameters in normal pancreas. World J Gastrointest Surg 2024; 16(7): 2031-2039
- URL: https://www.wjgnet.com/1948-9366/full/v16/i7/2031.htm
- DOI: https://dx.doi.org/10.4240/wjgs.v16.i7.2031