Retrospective Study
Copyright ©The Author(s) 2024.
World J Gastrointest Surg. Jul 27, 2024; 16(7): 2031-2039
Published online Jul 27, 2024. doi: 10.4240/wjgs.v16.i7.2031
Table 2 Intraclass correlation coefficient results of parameter values in different parts of the pancreas between different devices
Machine A vs machine B
Machine A vs machine C
Machine B vs machine C
ICC (95%CI)
P value
ICC (95%CI)
P value
ICC (95%CI)
P value
D in PaH0.274 (-0.731, 0.697)0.232 0.342 (-0.411, 0.712)0.145 0.728 (0.358, 0.886)0.002
D* in PaH-0.186 (-1.545, 0.482)0.670 0.27 (-0.611, 0.685)0.220 0.578 (0.02, 0.822)0.022
f in PaH-0.209 (-1.818, 0.491)0.673 0.173 (-0.386, 0.582)0.272 0.19 (-0.541, 0.622)0.274
D in PaB0.456 (-0.332, 0.776)0.089 0.683 (-0.026, 0.886)0.028 0.551 (-0.079, 0.814)0.037
D* in PaB-0.521 (-3.056, 0.393)0.815 -0.023 (-1.636, 0.587)0.519 -0.256 (-2.207, 0.491)0.691
f in PaB0.385 (-0.407, 0.739)0.124 -0.087 (-1.441, 0.534)0.580 -0.332 (-2.715, 0.515)0.748
D in PaT0.324 (-0.614, 0.718)0.186 0.286 (-0.557, 0.69)0.201 0.908 (0.781, 0.961)< 0.001
D* in PaT0.569 (-0.204, 0.836)0.060 -0.289 (-2.251, 0.475)0.713 -0.062 (-1.02, 0.504)0.569
f in PaT0.531 (-0.144, 0.806)0.047 0.35 (-0.633, 0.735)0.174 0.486 (-0.275, 0.789)0.073
SNR0.081 (-0.206, 0.421)0.3120.178 (-0.238, 0.551)0.2050.067 (-0.152, 0.364)0.301
ADC value0.870 (0.691, 0.946)< 0.0010.707 (0.393, 0.871)< 0.0010.808 (0.590, 0.917)< 0.001