All patients | Implantation date | |||||
2000-2011 (n = 56) | 2000-2007 (n = 37) | 2007-2011 (n = 19) | ||||
n (%) | Per 100 PY | n (%) | Per 100 PY | n (%) | Per 100 PY | |
Catheter inspection | 2 (4.0) | 0.7 | 2 (5.3) | 0.9 | 0 (0.0) | 0.0 |
Catheter replacement | 13 (26.0) | 4.6 | 8 (21.1) | 3.5 | 5 (41.7) | 8.8 |
Explantation of pump and catheter | 17 (34.0) | 6.0 | 151 (39.5) | 6.6 | 21 (16.7) | 3.5 |
Repositioning of pump | 2 (4.0) | 0.7 | 2 (5.3) | 0.9 | 0 (0.0) | 0.0 |
Fixation of pump | 2 (4.0) | 0.7 | 1 (2.6) | 0.4 | 1 (8.3) | 1.8 |
Cutaneous problem | 0 (0.0) | 0.0 | 0 (0.0) | 0.0 | 0 (0.0) | 0.0 |
Intra-abdominal problem | 1 (2.0) | 0.4 | 1 (2.6) | 0.4 | 0 (0.0) | 0.0 |
Remove clot at tip of catheter/flush catheter | 7 (14.0) | 2.5 | 4 (10.5) | 1.8 | 3 (25.0) | 5.3 |
Haematoma | 1 (2.0) | 0.4 | 1 (2.6) | 0.4 | 0 (0.0) | 0.0 |
Reposition of catheter | 3 (6.0) | 1.1 | 3 (7.9) | 1.3 | 0 (0.0) | 0.0 |
Infection | 2 (4.0) | 0.7 | 1 (2.6) | 1.2 | 1 (8.3) | 1.8 |
All | 50 (100.0) | 17.7 | 38 (100.0) | 44.3 | 12 (100.0) | 21.1 |
- Citation: Dijk PRV, Logtenberg SJ, Groenier KH, Haveman JW, Kleefstra N, Bilo HJ. Complications of continuous intraperitoneal insulin infusion with an implantable pump. World J Diabetes 2012; 3(8): 142-148