Copyright
©2012 Baishideng.
World J Diabetes. Aug 15, 2012; 3(8): 142-148
Published online Aug 15, 2012. doi: 10.4239/wjd.v3.i8.142
Published online Aug 15, 2012. doi: 10.4239/wjd.v3.i8.142
All patients | Implantation date | |||||
2000-2011 (n = 56) | 2000-2007 (n = 37) | 2007-2011 (n = 19) | ||||
n (%) | Per 100 PY | n (%) | Per 100 PY | n (%) | Per 100 PY | |
Haematoma | 3 (4.3) | 1.1 | 2 (3.8) | 0.9 | 1 (5.9) | 1.8 |
Infection | 7 (10.0) | 2.5 | 4 (7.5) | 1.8 | 3 (17.6) | 5.3 |
Pain | 11 (15.7) | 3.9 | 8 (15.1) | 3.5 | 3 (17.6) | 5.3 |
Cutaneous erosion | 2 (2.9) | 0.7 | 2 (3.8) | 0.9 | 0 (0.0) | 0.0 |
Dislocation | 3 (4.3) | 1.1 | 2 (3.8) | 0.9 | 1 (5.9) | 1.8 |
Hardware failure | 0 (0.0) | 0.0 | 0 (0.0) | 0.0 | 0 (0.0) | 0.0 |
Premature battery end of life | 0 (0.0) | 0.0 | 0 (0.0) | 0.0 | 0 (0.0) | 0.0 |
Insulin aggregate | 4 (5.7) | 1.4 | 4 (7.5) | 1.8 | 0 (0.0) | 0.0 |
Catheter occlusion | 23 (32.9) | 8.1 | 16 (30.2) | 7.1 | 7 (41.2) | 12.3 |
Encapsulation of the catheter tip | 3 (4.3) | 1.1 | 3 (5.7) | 1.3 | 0 (0.0) | 0.0 |
Peritonitis | 1 (1.4) | 0.4 | 1 (1.9) | 0.4 | 0 (0.0) | 0.0 |
Pump dysfunction | 12 (17.1) | 4.2 | 111 (20.8) | 4.9 | 11 (5.9) | 1.8 |
Other | 1 (1.4) | 0.4 | 0 (0.0) | 0.0 | 1 (5.9) | 1.8 |
All | 70 (100.0) | 24.8 | 53 (100.0) | 23.5 | 17 (100.0) | 29.9 |
- Citation: Dijk PRV, Logtenberg SJ, Groenier KH, Haveman JW, Kleefstra N, Bilo HJ. Complications of continuous intraperitoneal insulin infusion with an implantable pump. World J Diabetes 2012; 3(8): 142-148
- URL: https://www.wjgnet.com/1948-9358/full/v3/i8/142.htm
- DOI: https://dx.doi.org/10.4239/wjd.v3.i8.142