Minireviews
Copyright ©2013 Baishideng Publishing Group Co.
World J Gastrointest Endosc. Nov 16, 2013; 5(11): 540-550
Published online Nov 16, 2013. doi: 10.4253/wjge.v5.i11.540
Table 3 Comparison of different types of optical coherence tomography systems
ParametersTD-OCTSD-OCTSS-OCT/OFDI
MechanismInterference signals are detected as a function of optical time delay between obj. and ref. arm.Interference signals are detected with a camera as a function of optical frequencySpectral fringes are mapped to time domain by use of a swept laser and are measured with a detector as a function of time
Major componentsBroadband laser, optical delay line and a detectorBroadband laser, spectrometer and cameraTunable laser, digitizer and a balanced detector
Spectrum800 nm, 1000 nm, 1300 nm800 nm, 1000 nm, 1300 nm800 nm, 1000 nm, 1300 nm
Imaging depth1-3 mm1-3 mm1-3 mm
Resolution≥ 10 μm1-10 μm1-10 μm
Imaging speed (axial scan rate)Slow ( ≤ 5 kHz)Fast (20-150 kHz)Fairly fast (20-400 kHz)
SNRLowHighHigh
Image qualityModerateFairly highHigh
SensitivityLow (70-90 dB)High (85-105 dB)High (≥ 100 dB)
Phase stabilityLowHighModerate
PortabilityYesYesYes
System costLowHighModerate