Working channel findings | Control group (n = 30) | Experimental group (n = 30) | χ2 | P value |
Dark colored debris | 4 (13.3) | 2 (6.7) | 0.74 | 0.671 |
Light colored debris | 20 (66.7) | 13 (43.3) | 3.30 | 0.069 |
Residual debris | 22 (73.3) | 14 (46.7) | 4.44 | 0.035 |
Opaque fluid | 2 (6.7) | 3 (10.0) | 0.22 | 1.00 |
Clear fluid | 14 (46.7) | 18 (60.0) | 1.07 | 0.438 |
- Citation: Du J, Zhang M, Tao SY, Ye LS, Gong H, Hu B, Zhang QY, Qiao F. Efficacy of spray flushing in the reprocessing of flexible endoscopes: A randomized controlled trial. World J Gastroenterol 2024; 30(31): 3680-3688
- URL: https://www.wjgnet.com/1007-9327/full/v30/i31/3680.htm
- DOI: https://dx.doi.org/10.3748/wjg.v30.i31.3680