Table 1 Endoscopic working channel cleanliness scale
Working channel findings
Endoscope inspection ratings
Dark colored debris0 (none): No dark colored debris
1 (mild): Small dark particles or linear debris in up to 3 locations
2 (moderate): Small dark particles or debris in > 3 locations or larger debris in 1 location
3 (severe): Larger dark particles or debris in > 1 location
Light colored debris0 (none): No light colored/reflective debris
1 (mild): Small light-colored/reflective particles or residue in up to 3 locations
2 (moderate): Small light-colored/reflective particles or residue in > 3 locations, or larger debris/residue in 1
3 (severe): Larger light-colored/reflective particles or debris in > 1 location or present in continuous large
Opaque fluid0 (none): No opaque fluid
1 (mild): Up to 5 individual drops of opaque fluid within channel
2 (moderate): 6 to 10 individual drops of opaque fluid within channel
3 (severe): Greater than 10 individual drops of opaque fluid or pooling of fluid within channel
Clear fluid0 (none): No clear fluid
1 (mild): Up to 5 individual drops of clear fluid within channel
2 (moderate): 6 to 10 individual drops of clear fluid within channel
3 (severe): Greater than 10 individual drops of clear fluid or pooling of fluid within channel