Working channel findings | Endoscope inspection ratings |
Dark colored debris | 0 (none): No dark colored debris |
1 (mild): Small dark particles or linear debris in up to 3 locations | |
2 (moderate): Small dark particles or debris in > 3 locations or larger debris in 1 location | |
3 (severe): Larger dark particles or debris in > 1 location | |
Light colored debris | 0 (none): No light colored/reflective debris |
1 (mild): Small light-colored/reflective particles or residue in up to 3 locations | |
2 (moderate): Small light-colored/reflective particles or residue in > 3 locations, or larger debris/residue in 1 | |
3 (severe): Larger light-colored/reflective particles or debris in > 1 location or present in continuous large | |
Opaque fluid | 0 (none): No opaque fluid |
1 (mild): Up to 5 individual drops of opaque fluid within channel | |
2 (moderate): 6 to 10 individual drops of opaque fluid within channel | |
3 (severe): Greater than 10 individual drops of opaque fluid or pooling of fluid within channel | |
Clear fluid | 0 (none): No clear fluid |
1 (mild): Up to 5 individual drops of clear fluid within channel | |
2 (moderate): 6 to 10 individual drops of clear fluid within channel | |
3 (severe): Greater than 10 individual drops of clear fluid or pooling of fluid within channel |