Cap | Endocuff | Endocuff Vision | Endorings | |
Manufacturer | Olympus, Centre Valley, Pennsylvania | Arc Medical Leeds, United Kingdom | Norgine Pharmaceuticals Ltd, Uxbridge, United Kingdom | EndoAid, Caesarea, Israel |
Launched in market | 1993 | 2011 | 2016 | 2015 |
Short description | Transparent, single-use distal attachment with side hole for draining of fluid | Single-use, soft, radiopaque, 2 cm long cylindrical sleeve with flexible projections arranged in 2 rows of 8, emerging from gaps on the shaft of the device | Single-use, device with single row of 8 flexible 15 mm spikes | Single-use device composed of 2 layers of flexible, soft circular rings, placed on a cylindrical cuff |
Material | Thermoplastic elastomer | Core: Non-latex, biocompatible polymer; Projections: thermoplastic elastomer | Latex free, polypropylene | Silicone |
Dimensions | Outer diameter ranging from 13.9-16.1 mm according to each type of cap | Finger projections: proximal 8.15 mm, distal 5mm; core length: 23.8 mm; diameter: 16.1, 16.7, 17.2, and 18.5 mm (hairs folded back) and 32.6, 33.1, 33.6, and 34.8 mm (hairs opened out) | Diameter: 16.1, 16.7, 17.2, and 18.5mm (spikes folded back) and 39.07, 39.07, 39.07, and 39.66 mm (spikes opened out) | 22-50 mm diameter |
Mode of action | Protruding cap manipulates and flattens haustral folds to inspect the mucosa on the proximal side of the fold maintaining optimal field of view | Hinged projections flatten and spread mucosa and folds | Hinged projections flatten and improve visibility behind the colon folds | Sequential rings stretches out the folds of the colon during withdrawal for a clear view |
Interfere with view of field | Edge of the hood comes into the vision field of the colonoscope, but lesions can be seen through the transparent wall | No interference of vision | No interference with vision | No interference with vision |
Compatible scopes | Adult, pediatric: Ten different sizes, to fit all scopes | Adult, pediatric: 4 color-coded sizes (purple, orange, green and blue) to fit all scopes | Adult, pediatric: 4 color-coded sizes (purple, orange, green and blue) to fit all scopes | Scope Distal End Diameter [mm]; Adult colonoscope 12.8-14.5 mm; Slim Adult colonoscope 11.5-13.0 mm |
Advantages | Resection of wider areas; Suction and insufflation of air unaffected | Folds movement provides a dynamic picture - even the smallest polyps can be identified; Centers the scope in the middle of the lumen preventing sudden slip back and “red-out”; Projections allow traction to avoid sudden slippage around turns and flexures, improving scope’s stability; Helps perform EMR | Delivers more tip control without compromising intubation - improving loop management; Early and controlled view of the upstream surface of large folds - no need for repeated intubation; Prevents sudden slip back and red out; Optimizes tip position during therapy and polyp retrieval | Maintains position during loop reduction, decreases slippage, anchoring during endoscopic therapy; Maintains identical depth and breadth of scope's viewing field; Minimal resistance on insertion; Easy ileum intubation |
Disadvantages | Interfere with the field of view | Petechial marks on colon; Potential dislodgement; Larger model more effective than smaller; Ileum intubation may be difficult | Potential dislodgement | Ileum intubation may be difficult |